List of identified Argon lines in offered spectroscopic modes

Fitting information

Mode Order Fit NOL Residual Dispersion EWR
S27_3_SK 2 3 13 0.48 4.97 2.02 - 2.53
S54_2_SK 2 4 18 0.62 9.75 1.79 - 2.49
S54_3_SK 2 4 23 0.76 10.01 1.79 - 2.57
S27_2_SK 2 4 15 0.33 4.86 1.79 - 2.24
S54_4_SK 1 4 15 1.30 19.56 1.79 - 2.57
S54_3_H 3 3 16 0.15 6.81 1.50 - 1.84 (1)
S54_3_SH 3 3 23 0.37 6.85 1.37 - 1.84
S27_3_SH 3 4 21 0.28 3.41 1.37 - 1.72
S27_4_SH 1 4 17 0.93 9.67 1.37 - 1.84
S54_4_SJ 1 3 12 1.18 20.04 0.91 - 1.40 (2)
S27_4_SJ 1 3 10 0.77 9.92 1.26 - 1.40 (3)
S54_4_SJ 2 3 13 0.80 9.78 0.91 - 1.37
S27_4_SJ 2 3 6 0.33 4.85 0.91 - 1.14


EWR = Effective wavelength range.Includes filter, atmosphere and order overlap.
NOL = number of lines used in the fit. Faint or blended lines were excluded.
(1) 2nd order lines from 1.75 to 1.85 microns
(2) 2nd order lines from 0.91 to 1.37 microns
(3) 2nd order lines from 0.91 to 1.11 micronsNote: Blended at LR means that the line is blended with some other lineat the resolution of the grism with the 86mas slit.

List of identified lines in offered spectroscopic modes

Spect. mode Wavelength Strength Comments - Spect. mode Wavelength Strength Comments
S27_3_SK 20322.56 380000   - S54_2_SK 17919.61 230000  
  20574.43 50000 In the wings of 20621.86 |   18432.80 1400000  
  20621.86 5000000   |   18491.76 50000 In the wings of 18432.80
  20652.77 140000 In the wings of 20621.86 |   18569.30 - Blended with 18575.23
  20739.22 200000   |   18575.23 - Blended with 18569.30
  20816.72 150000   |   18637.37 110000  
  20991.84 3000000   |   18750.12 60000  
  21041.57 30000 In the wings of 20991.84 |   19822.91 700000  
  21338.71 300000   |   19971.18 750000  
  21540.09 1200000   |   20031.14 - Blended with 20035.57
  22045.58 300000   |   20035.57 - Blended with 20031.14
  22083.21 1300000   |   20074.41 50000  
  22118.66 50000 In the wings of 22083.21 |   20322.56 420000  
  23139.52 1000000   |   20574.43 - In the wings of 20621.86
  23851.54 850000   |   20621.86 5300000  
  23973.06 850000   |   20739.22 170000  
  23132.12 300000   |   20816.72 150000  
- - - - |   20991.84 3300000  
        |   21041.57 - In the wings of 20991.84
        |   21338.71 320000  
        |   21540.09 1300000  
        |   22045.58 300000 In the wings of 22083.21
        |   22083.21 1300000  
        |   23139.52 1000000  
        |   23851.54 900000  
        |   23973.06 900000  
- - - - - - - - -
S54_3_SK 17919.61 210000   | S54_4_SK 17919.61 210000  
  18432.80 1000000   |   18432.80 1000000  
  18491.76 50000 In the wings of 18432.80 |   18491.76 - In the wings of 18432.80
  18569.30 - Blended with 18575.23 |   18569.30 - Blended with 18575.23
  18575.23 - Blended with 18569.30 |   18575.23 - Blended with 18569.30
  18637.37 75000   |   18637.37 75000  
  18750.12 45000   |   18750.12 45000  
  19822.91 480000   |   19822.91 480000  
  19971.18 500000   |   19971.18 500000  
  20031.14 - Blended with 20035.57 |   20031.14 - Blended with 20035.57
  20035.57 - Blended with 20031.14 |   20035.57 - Blended with 20031.14
  20074.41 30000   |   20322.56 250000  
  20322.56 250000   |   20574.43 - In the wings of 20621.86
  20574.43 - In the wings of 20621.86 |   20621.86 3000000  
  20621.86 3000000   |   20739.22 100000  
  20739.22 100000   |   20816.72 90000  
  20816.72 90000   |   20991.84 2000000  
  20991.84 2000000   |   21041.57 - In the wings of 20991.84
  21041.57 - In the wings of 20991.84 |   21338.71 200000  
  21338.71 200000   |   21540.09 800000  
  21540.09 800000   |   22045.58 - In the wings of 22083.21
  22045.58 200000 In the wings of 22083.21 |   22083.21 1000000  
  22083.21 1000000   |   23139.52 700000  
  23139.52 700000   |   23851.54 600000  
  23851.54 600000   |   23973.06 630000  
  23973.06 630000   |   25132.12 240000  
  25132.12 240000   |   25512.18 170000  
  25512.18 170000   |   25668.0 80000  
  25668.0 80000   | - - - -
- - - - - - - - -
S27_4_SK 17919.61 210000   - S27_2_SK 17919.61 210000  
  18432.80 1000000   |   18432.80 1000000  
  18491.76 50000 In the wings of 18432.80 |   18491.76 50000 In the wings of 18432.80
  18569.30 - Blended with 18575.23 |   18569.30 - Blended with 18575.23
  18575.23 - Blended with 18569.30 |   18575.23 - Blended with 18569.30
  18637.37 75000   |   18637.37 75000  
  18750.12 45000   |   18750.12 45000  
  19822.91 480000   |   19822.91 480000  
  19971.18 500000   |   19971.18 500000  
  20031.14 - Blended with 20035.57 |   20031.14 - Blended with 20035.57
  20035.57 - Blended with 20031.14 |   20035.57 - Blended with 20031.14
  20074.41 30000   |   20074.41 30000  
  20322.56 250000   |   20322.56 250000  
  20574.43 - In the wings of 20621.86 |   20574.43 - In the wings of 20621.86
  20621.86 3000000   |   20621.86 3000000  
  20652.77 - In the wings of 20621.86 |   20652.77 - In the wings of 20621.86
  20739.22 100000   |   20739.22 100000  
  20816.72 90000   |   20816.72 90000  
  20991.84 2000000   |   20991.84 2000000  
  21041.57 - In the wings of 20991.84 |   21041.57 - In the wings of 20991.84
  21338.71 200000   |   21338.71 200000  
  21540.09 800000   |   21540.09 800000  
  22045.58 200000   |   22045.58 - In the wings of 22083.21
  22083.21 1000000   |   22083.21 1000000  
- - - - - - - - -
S54_3_H 14600.46 40000   - S54_3_SH 13722.3 270000  
  14638.42 - Blended with 14654.35 |   13914.4 100000  
  14654.35 - Blended with 14638.42 |   14097.9 2300000  
  14697.03 40000   |   14600.46 270000  
  14743.17 100000   |   14638.42 - Blended with 14654.35
  15050.62 4000000   |   14654.35 - Blended with 14638.42
  15176.84 1000000   |   14697.03 100000  
  15306.07 1200000   |   14743.17 170000  
  15333.54 550000   |   15050.62 2200000  
  15353.44 - Blended with 15357.32 |   15176.84 600000  
  15357.32 - Blended with 15353.44 |   15306.07 640000  
  15406.85 230000   |   15333.54 420000  
  15450.95 20000   |   15353.44 - Blended with 15357.32
  15559.71 25000   |   15357.32 - Blended with 15353.44
  15739.21 30000   |   15406.85 130000  
  15821.10 25000   |   15450.95 8000  
  15904.03 330000 Blended with 15887.51 |   15559.71 15000  
  15993.86 1000000   |   15739.21 14000  
  16127.06 120000   |   15821.10 25000  
  16184.44 170000   |   15904.03 190000 Blended with 15887.51
  16268.51 25000   |   15993.86 600000  
  16441.07 330000   |   16127.06 75000  
  16524.38 900000   |   16184.44 100000  
  16553.78 170000   |   16268.51 15000  
  16744.65 280000   |   16441.07 180000  
  16945.21 5000000   |   16524.38 500000  
  17449.67 260000   |   16553.78 110000  
  17828.87 50000   |   16744.65 160000  
  17919.61 1000000   |   16945.21 2900000  
- - - - |   17449.67 150000  
        |   17828.87 28000  
        |   17919.61 600000  
        |   18432.8 46000  
- - - - - - - - -
S27_3_SH 13914.4 100000   - S27_4_SH 13722.3 270000  
  14097.9 2300000   |   13914.4 100000  
  14600.46 270000   |   14097.9 2300000  
  14638.42 390000   |   14600.46 - In the wing of 14638.42
  14654.35 410000   |   14638.42 - Blended with 14654.35
  14697.03 100000   |   14654.35 - Blended with 14638.42
  14743.17 170000   |   14697.03 100000  
  15050.62 2200000   |   14743.17 170000  
  15176.84 600000   |   15050.62 2200000  
  15306.07 640000   |   15176.84 600000  
  15333.54 420000   |   15306.07 - Blended with several lines
  15353.44 - Blended with 15357.32 |   15333.54 - Blended with several lines
  15357.32 - Blended with 15353.44 |   15353.44 - Blended with several lines
  15406.85 130000   |   15357.32 - Blended with several lines
  15450.95 8000   |   15406.85 130000  
  15559.71 15000   |   15559.71 15000  
  15739.21 14000   |   15739.21 14000  
  15821.10 25000   |   15821.10 25000  
  15904.03 190000 Blended with 15887.51 |   15904.03 190000 Blended with 15887.51
  15993.86 600000   |   15993.86 600000  
  16127.06 75000   |   16127.06 75000  
  16184.44 100000   |   16184.44 100000  
  16268.51 15000   |   16268.51 15000  
  16441.07 180000   |   16441.07 180000  
  16524.38 500000   |   16524.38 500000  
  16553.78 110000   |   16553.78 110000  
  16744.65 160000   |   16744.65 160000  
  16945.21 2900000   |   16945.21 2900000  
- - - - |   17449.67 150000  
        |   17828.87 28000  
        |   17919.61 600000  
        |   18432.8 46000  
- - - - - - - - -
S54_3_SJ 9125.47 2400000 (see notes) | S27_4_SJ 12406.22 1400000 (see notes)
  9227.03 940000   |   12491.08 2000000  
  9356.78 110000   |   12806.20 900000  
  9660.43 4000000   |   12960.20 2900000  
  9787.18 840000   |   13011.82 1000000  
  10472.92 4200000   |   13276.27 2300000  
  10676.49 1500000   |   13316.85 2000000  
  11081.90 280000   |   13370.77 2700000  
  11491.25 2600000 Blended with 11444.97 |   13507.88 4500000  
  11671.90 1300000   |   13626.38 1600000  
  12705.75 4000000 Blended with 12806.24 |   13722.30 5800000  
  12806.24 4000000 Blended with 12705.75 | - - - -
  12960.20 1.4e7 Blended with 12936.73 and 13011.82 |        
  13370.77 1.5e7   |        
  13507.88 2.0e7   |        
  13626.38 1.3e7   |        
  13722.3 2.9e7   |        
- - - - - - - - -

 


Notes: Many heavily blended lines. Only the cleanest lines are reported here.- See the ISAAC list for additional lines.