back to: | dfos | tqs |
|
||||||||||||||||||
|
||||||||||||||||||
NAME | qc_HC_AO.sh | |||||||||||||||||
VERSION |
1.0 -- 2006-06-27 fill template by handbook 1.1 -- 2006-07-10 reviseted | |||||||||||||||||
INSTRUMENT | SINFONI | |||||||||||||||||
SYNTAX | SHELL | |||||||||||||||||
RAWTYPE | HC_AO | |||||||||||||||||
CALL |
measureQuality from $DFS_PRODUCT/HC_AO/$DATE: qc_HC_AO -d $DATE -f $primfile | |||||||||||||||||
PURPOSE |
a) generates QC report, b) writes QC1 parameters into local database Two templaes in operation, one with one camera, and one with two cameras | |||||||||||||||||
PROCINPUT |
a) $DATE in yyyy-mm-dd, b) $primfile is the PRO.CATG=MASTER_PSF product with the _0000.fits extension. c) raw frames are implicitely read from the AB. | |||||||||||||||||
QC1TABLE | trending | table in QC1 database: sinfoni_psf | |||||||||||||||||
TRENDPLOT | trending | current trending plot associated to this procedure: trend_SI_HC_AO_STREHL_S_current.gif trend_SI_HC_AO_STREHL_M_current.gif | |||||||||||||||||
QC1PAGE | trending | associated trending page: qc_HC_AO.html | |||||||||||||||||
QC1PLOTS |
hcao0.png display STACKED intermediate product on Real Time Display hcao1.png display MASTER_PSF product on Real Time Display hcao2.png UL: Extracted spectrum: (red, black, blue). 2L: centroid position of the spectrum (x and y). 3L: Position angle of a 2d-Gaussian fit. LL FWHM x and y and FWHM ratio of the spectrum. UR encircled energy as a function of radius from collpsed frame 2R Strehl ratio as a function of binned wavelength intervalls | |||||||||||||||||
QC1PARAM |
QC1 parameters, read from the RealTimeComputer of the AO module (as for PSF, OBSNOD ...) fits header key name , databse key name : AOS.RTC.GUIDESTAR.MAGNITUDE , TBD AOS.RTC.OBS.FWHMMEAN , TBD AOS.RTC.OBS.FWHMSTD , TBD AOS.RTC.OBS.ENCMEAN , TBD AOS.RTC.OBS.ENCSTD , TBD AOS.RTC.OBS.STREHLMEAN , TBD NONE , qc_oes1, oddeven column stripe of first SKY frame, from post-pipeline QC script QC1 parameters, generated by a post-pipeline QC script that does a Gaussian fit to all planes of the final cube: X_POS , TBD Y_POS , TBD FWHM_MIN , TBD FWHM_MAX , TBD FWHM_POSANG , TBD FWHM_RATIO , TBD TOT , TBD QC1 parameters generated by the pipeline (a standard sequence) QC.FRMON.MEANFLUX, qc_frmon_meanflux , standard QC1 parameter: mean flux of lamp=on frame QC.FRMOFF.MEANFLUX, qc_frmoff_meanflux, standard QC1 parameter: mean flux of lamp=off frame QC.FRMDIF.MEANFLUX, qc_frmdif_meanflux standard QC1 parameter: mean flux of lamp=on frame minus lamp=off frame QC.FRMON.MAXFLUX, qc_frmon_maxflux standart QC1 parameter: max flux of lamp=on frame QC.FRMOFF.MAXFLUX, qc_frmoff_maxflux standart QC1 parameter: max flux of lamp=off frame QC.FRMDIF.MAXFLUX, qc_frmdif_maxflux standart QC1 parameter: max flux of lamp=on farme minus lam=off frame QC.FRMON.NPIXSAT , qc_frmon_npixsat Number of saturated pixels QC1 parameters generated by the pipeline for HC_AO only QC.FWHMX , qc_fwhmx QC.FWHMY , qc_fwhmy INS1.LAMP5.ST , ins_lamp5_st INS1.SHUT2.ST , ins_stut2_std QC.STREHL.MED , qc_strehl_ned QC.STREHL.MEDERR , qc_strehl_mederr | |||||||||||||||||
ALGORITHM |
Description of algorithms: (status 2006-07-10) RTC QC parameters : special AO software FWHM parameters : the post-pipeline scripts applied the ESO-MIDAS command CENTER/IQE to each plane of the final product cube. The wavelength mean MEDIAN value is given. FRM QC1 parameters : simple statistics on the final product cube FWHM parameters : Gauss fit integrated into the pipeline; QC1 parameter inventory TBD | |||||||||||||||||
CERTIF |
Reasons for rejection: none | |||||||||||||||||
COMMENTS | This is a daily health check calibration using the intrenal FIBER | |||||||||||||||||
created by qcDocu, a tqs tool |