NAME qc_HC_AO.sh
VERSION 1.0 -- 2006-06-27 fill template by handbook
1.1 -- 2006-07-10 reviseted
INSTRUMENT SINFONI
SYNTAX SHELL
RAWTYPE HC_AO
CALL measureQuality
from $DFS_PRODUCT/HC_AO/$DATE:
qc_HC_AO -d $DATE -f $primfile
PURPOSE a) generates QC report,
b) writes QC1 parameters into local database
Two templaes in operation, one with one camera, and one with two cameras
PROCINPUT a) $DATE in yyyy-mm-dd,
b) $primfile is the PRO.CATG=MASTER_PSF product with the _0000.fits extension.
c) raw frames are implicitely read from the AB.
QC1TABLE trending | table in QC1 database: sinfoni_psf
TRENDPLOT trending | current trending plot associated to this procedure: trend_SI_HC_AO_STREHL_S_current.gif trend_SI_HC_AO_STREHL_M_current.gif
QC1PAGE trending | associated trending page: qc_HC_AO.html
QC1PLOTS

hcao0.png
display STACKED intermediate product on Real Time Display
hcao1.png
display MASTER_PSF product on Real Time Display
hcao2.png
UL: Extracted spectrum: (red, black, blue).
2L: centroid position of the spectrum (x and y).
3L: Position angle of a 2d-Gaussian fit.
LL FWHM x and y and FWHM ratio of the spectrum.
UR encircled energy as a function of radius from collpsed frame
2R Strehl ratio as a function of binned wavelength intervalls
QC1PARAM QC1 parameters, read from the RealTimeComputer of the AO module (as for PSF, OBSNOD ...)
fits header key name , databse key name : AOS.RTC.GUIDESTAR.MAGNITUDE , TBD
AOS.RTC.OBS.FWHMMEAN , TBD
AOS.RTC.OBS.FWHMSTD , TBD
AOS.RTC.OBS.ENCMEAN , TBD
AOS.RTC.OBS.ENCSTD , TBD
AOS.RTC.OBS.STREHLMEAN , TBD

NONE , qc_oes1, oddeven column stripe of first SKY frame, from post-pipeline QC script

QC1 parameters, generated by a post-pipeline QC script that does a Gaussian fit to all planes of the final cube:
X_POS , TBD
Y_POS , TBD
FWHM_MIN , TBD
FWHM_MAX , TBD
FWHM_POSANG , TBD
FWHM_RATIO , TBD
TOT , TBD

QC1 parameters generated by the pipeline (a standard sequence)
QC.FRMON.MEANFLUX, qc_frmon_meanflux , standard QC1 parameter: mean flux of lamp=on frame
QC.FRMOFF.MEANFLUX, qc_frmoff_meanflux, standard QC1 parameter: mean flux of lamp=off frame
QC.FRMDIF.MEANFLUX, qc_frmdif_meanflux standard QC1 parameter: mean flux of lamp=on frame minus lamp=off frame
QC.FRMON.MAXFLUX, qc_frmon_maxflux standart QC1 parameter: max flux of lamp=on frame
QC.FRMOFF.MAXFLUX, qc_frmoff_maxflux standart QC1 parameter: max flux of lamp=off frame
QC.FRMDIF.MAXFLUX, qc_frmdif_maxflux standart QC1 parameter: max flux of lamp=on farme minus lam=off frame
QC.FRMON.NPIXSAT , qc_frmon_npixsat Number of saturated pixels

QC1 parameters generated by the pipeline for HC_AO only
QC.FWHMX , qc_fwhmx
QC.FWHMY , qc_fwhmy
INS1.LAMP5.ST , ins_lamp5_st
INS1.SHUT2.ST , ins_stut2_std
QC.STREHL.MED , qc_strehl_ned
QC.STREHL.MEDERR , qc_strehl_mederr
ALGORITHM Description of algorithms: (status 2006-07-10)
RTC QC parameters : special AO software
FWHM parameters : the post-pipeline scripts applied the ESO-MIDAS command CENTER/IQE to each plane of the final product cube. The wavelength mean MEDIAN value is given.
FRM QC1 parameters : simple statistics on the final product cube
FWHM parameters : Gauss fit integrated into the pipeline; QC1 parameter inventory TBD
CERTIF Reasons for rejection: none
COMMENTS This is a daily health check calibration using the intrenal FIBER
created by qcDocu, a tqs tool