WFI N2 Flushing System
The WFI N2 Flushing System There is a solenoid valve mounted on the telescope base at the low pressure side of the N2 gas supply. This valve is driven by the TCS VME computer under the supervision of a program task plus a set of commands that can be issued from the vxte2p2 console (or from any console using an rlogin) for; opening the valve, closing the valve, reading the status, and switching the system to automatic and manual modes.
The commands are the following five:
- N2_status
- N2_mode_manual
- N2_mode_automatic
- Checks if the system is in manual or automatic mode.
- If in manual mode; send the above mentioned warning message to the console every 5 seconds.
- If in automatic mode; read the humidity every 2 seconds, average 150 values and open the valve if the humidity average is >= 60% and close the valve if the humidity average is <= 50%.
- N2_open
- N2_close
Note1: Always at boot time the conditions by default are the following:
- System mode in automatic.
- <=50% closing and >=60% opening.
Note2: After boot it is possible, if desired, to change the opening and closing points just by changing the globals, "double N2cls = X.X" and "double N2opn = X.X" to any value between 0.0 to 100.0.