VLT Environmental Specifications

Telescope operating temperature range (full performance) 0°C to 15°C
Functional temperature range -10°C to 30°C
Coudé lab operating range -5°C to 20°C
Dust environment in enclosures and Coudé labs Class 30000
Wind:
»Operation, enclosure open <14 m/s (into wind, 20 min. average) Full performance
<18 m/s (elsewhere, 20 min. average) Full performance
<27 m/s (gust) Full performance
»Survival, enclosure open 36 m/s (gust) Minor system damage (repair time<3 weeks)
»Survival, enclosure closed 47.2 m/s (gust) Minor system damage (repair time<3 weeks)
54.2 m/s (gust) Major system damage(repair time>3 weeks)
Maximum humidity during observation 95 % RH
Earthquakes:
»Normal operation or standby Moderate earthquakes Negligible system damage
OBE Minor system damage (repair time<3 weeks)
MLE Major system damage(repair time>3 weeks)
»Maintenance operations Relaxed requirements