This documentation is intended both for QC scientists and SciOps astronomers (who may want to ignore the technical information displayed in grey).
|
|
|
|
|
bottom
plots
|
contact
|
|
NAME |
espre_wave.py
|
|
VERSION |
1.0 -- October 2018
1.0.1 -- new QC parameters: delta_time_fp, pipe_order_min/max (2018-11-15)
1.0.2 -- new QC parameters: fib[a,b]_counts_max[blux,red] for LFC (2019-07-16)
1.0.3 -- changes for espdr-1.3.0 (2019-07-18)
1.0.4 -- proc_vers QC1 parameter added (2020-08-14)
1.0.5 -- change of parameters in espre_wavelines (2018-08-20)
|
|
SYNTAX |
PYTHON
|
|
CALL |
processQC
from $DFS_PRODUCT/WAVE__/$DATE
/home/python/espre_wave.py -i -a $AB
|
|
INSTRUMENT |
ESPRESSO
|
|
RAWTYPE |
WAVE__
|
|
PURPOSE |
QC check of raw and product wavelength calibrations
|
|
PROCINPUT |
raw and product wavlength calibration frames
|
|
QC1TABLE |
trending | table(s) in QC1 database:
espresso_wave_thar_a
|
|
TRENDPLOT |
trending | HealthCheck plot(s) associated to this procedure:
trend_report_WAVE_FP_HR1x1_HC.html |
trend_report_WAVELINES_HR1x1_HC.html |
trend_report_WAVE_THAR_HR1x1_HC.html |
trend_report_WAVEDISP_HR1x1_HC.html |
trend_report_WAVERESO_HR_HC.html |
trend_report_WAVE_LFC_HR1x1_HC.html |
trend_report_WAVE_DRIFT_HR1x1_HC.html
|
|
QC1PAGE |
trending | associated documentation:
waveFP_qc1.html
|
|
QC1PLOTS
top |
wave1.png
Example for WAVE_THAR_FP
left: used ThAr lines for each order
right: used FP lines for each order
wave2.png
left: resolving power R for each ThAr line
right: number of FP lines, chi^2 of wavelength solution, and RMS of fit for each order
wave3.png
upper left: cut in X direction through raw frame, lower Y position
upper middle: cut in X direction through raw frame, upper Y position
upper right: histogram of raw frame
upper left: cut in X direction through raw frame, lower Y position, detail
upper middle: cut in X direction through raw frame, upper Y position, detail
upper right: image of raw frame
wave4.png
extracted line spectra per order
|
|
QC1PARAM |
QC1 parameters written in QC1 database:
QC1 DB names: e.g. fiba_thar_lines_tot | fiba_disp_avg | ...
|
|
ALGORITHM |
Most pipeline QC parameters come per order.
For the QC1 database, either average/RMS/total values are calculated or values refer to the central order.
|
|
CERTIF |
Possilbe reasons for rejection: low line flux level or saturation.
|
|
COMMENTS |
None
|
top |
[2020-09-01T15:48:48] created by qcDocu v1.1.2, a tqs tool
|