mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
UVES:
wavelength calib derotator
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
UVES trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-11-21T13:23:31 (UT)
  | now: 2024-11-21T17:13:51 (UT) muc02 QC pipeline: uves-6.1.11 (installed 2023-06-22)
same group: Drot_resolution Drot_ThAR_lamp
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
UVES news:  
Report news:
DATE*: [?] 2024-11-14
report | NLT
2024-11-15
report | NLT
2024-11-16
report | NLT
2024-11-17
report | NLT
2024-11-18
report | NLT
2024-11-19
report | NLT
2024-11-20
report | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw raw
P...
CAL product quality:
products products products products products products products
 
transfer   ngas [?]

scores&comments | FULL | history ... | plot tutorial ... | contact | HC_docu... | quality events...               [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 60 NONE none  122 resol_med this | last_yr | all Average resolving power, BLUE 390.0, 1x1, 0.8arcs slit
2 QC1DB MEDIAN 57.6 NONE none  122 resol_med this | last_yr | all Average resolving power, REDU 580.0, 1x1, 0.8arcs slit
3 QC1DB MEDIAN 60.1 NONE none  122 resol_med this | last_yr | all Average resolving power, REDL 580.0, 1x1, 0.8arcs slit
4 QC1DB MEDIAN 2.81 Ang none  122 resol_rms this | last_yr | all RMS of resolving power BLUE, 390, 0.8
5 QC1DB MEDIAN 4.49 Ang none  122 resol_rms this | last_yr | all RMS of resolving power REDU, 580, 0.8
6 QC1DB MEDIAN 3.47 Ang none  122 resol_rms this | last_yr | all RMS of resolving power REDL, 580, 0.8
7 QC1DB MEDIAN 3.18 pix none  122 fwhm_med this | last_yr | all median FWHM of selected lines BLUE, 390, 0.8
8 QC1DB MEDIAN 4.36 pix none  122 fwhm_med this | last_yr | all median FWHM of selected lines REDU, 580, 0.8
9 QC1DB MEDIAN 4.18 pix none  122 fwhm_med this | last_yr | all median FWHM of selected lines REDL, 580, 0.8
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:resol_med
median:60NONE
thresholds:none 
N_data plotted:122
[click plot for closeup]
Plot 2
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:resol_med
median:57.6NONE
thresholds:none 
N_data plotted:122
[click plot for closeup]
Plot 3
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:resol_med
median:60.1NONE
thresholds:none 
N_data plotted:122
[click plot for closeup]
Plot 4
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:resol_rms
median:2.81Ang
thresholds:none 
N_data plotted:122
[click plot for closeup]
Plot 5
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:resol_rms
median:4.49Ang
thresholds:none 
N_data plotted:122
[click plot for closeup]
Plot 6
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:resol_rms
median:3.47Ang
thresholds:none 
N_data plotted:122
[click plot for closeup]
Plot 7
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:fwhm_med
median:3.18pix
thresholds:none 
N_data plotted:122
[click plot for closeup]
Plot 8
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:fwhm_med
median:4.36pix
thresholds:none 
N_data plotted:122
[click plot for closeup]
Plot 9
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:fwhm_med
median:4.18pix
thresholds:none 
N_data plotted:122
[click plot for closeup]
This plot

The Derotator trending plots are monitoring UVES Resolution based on QC parameters derived from Wavelength Calibration frames taken with the derotator moving in steps of 30 deg.
This set of calibrations is taken each 2 weeks and are used to check if changes in the whole optical path.
The data are taken at 390 and 580 with a slit of 0.8 arcs.

Drot_resolution:

  • plots 1-3: resolution = f (ang of the derotator)
  • plots 4-6: rms_resolution = f (ang of the derotator)
  • plots 7-9: median of the FWHM of the selected lines

    Drot_ThAr_lamp:
  • plots 1-3: resolution = f (ang of the derotator)
  • plots 4-6: intensity = f (ang of the derotator), this is the averaged intensity of the lines used in the solution
  • plots 7-9: nlines = f (ang of the derotator), number of lines used in the
  • plots 7-9: resid_rms = f (ang of the derotator), sigma of residual of line positions to fit solution The ThAr lines are identified by the pipeline and their average flux (intensity, HC Drot_ThAr_lamp) is calculated, lines with sufficient intensity (nlines, HC Drot_ThAr_lamp) are selected for polynomial fit Find more information about UVES Wavelength Solution monitoring here.  
    General information

    Click on any of the plots to see a close-up version.

    The latest date is indicated on top of the plot.

    If configured,

    • data points belonging to the latest date are specially marked
    • statistical averages are indicated by a solid line, and thresholds by broken lines
    • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
    • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
    • you can download the data for each parameter set if the 'Data downloads' link shows up