UVES trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-12-22T13:39:34 (UT)
  | now: 2024-12-22T20:12:59 (UT) muc02 QC pipeline: uves-6.1.11 (installed 2023-06-22)
same group: Res_BLUE Res_REDU Res_REDL   BLUE_CD1 B_CD2 REDL_CD3 RL_CD4 REDU_CD3 RU_CD4
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
UVES news:  
Report news:
DATE*: [?] 2024-12-15
report | NLT
2024-12-16
report | NLT
2024-12-17
report | NLT
2024-12-18
report | NLT
2024-12-19
report | NLT
2024-12-20
report | NLT
2024-12-21
report | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw raw
P...
CAL product quality:
products products products products products products products
 
transfer   ngas [?]

scores&comments | FULL | history ... | plot tutorial ... | contact | HC_docu... | quality events...                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 89.2 NONE none  14 resol_med this | last_yr | all Mean resolving power, BLUE 346.0nm 1x1 0.4arcs slit
1 QC1DB MEDIAN 49.9 NONE none  56 resol_med this | last_yr | all Mean resolving power, BLUE 346.0nm 1x1 1.0arcs slit
2 QC1DB MEDIAN 87.3 NONE none  204 resol_med this | last_yr | all Mean resolving power, BLUE 390.0nm 1x1 0.4arcs slit
2 QC1DB MEDIAN   NONE none  0 resol_med this | last_yr | all Mean resolving power, BLUE 390.0nm 1x1 0.6arcs slit
2 QC1DB MEDIAN 50.6 NONE none  175 resol_med this | last_yr | all Mean fwhm, BLUE 390.0nm 1x1 1.0arcs slit
3 QC1DB MEDIAN 86.8 NONE none  421 resol_med this | last_yr | all Mean resolving power, BLUE 437.0nm 1x1 0.4arcs slit
3 QC1DB MEDIAN 50.6 NONE none  103 resol_med this | last_yr | all Mean resolving power, BLUE 437.0nm 1x1 1.0arcs slit
4 QC1DB MEDIAN 2.11 NONE none  13 fwhm_med this | last_yr | all Mean fwhm, BLUE 346.0nm 1x1 0.4arcs slit
4 QC1DB MEDIAN 3.83 NONE none  55 fwhm_med this | last_yr | all Mean fwhm, BLUE 346.0nm 1x1 1.0arcs slit
5 QC1DB MEDIAN 2.17 NONE none  194 fwhm_med this | last_yr | all Mean fwhm, BLUE 390.0nm 1x1 0.4arcs slit
5 QC1DB MEDIAN   NONE none  0 fwhm_med this | last_yr | all Mean fwhm, BLUE 390.0nm 1x1 0.6arcs slit
5 QC1DB MEDIAN 3.77 NONE none  169 fwhm_med this | last_yr | all Mean fwhm, BLUE 390.0nm 1x1 1.0arcs slit
6 QC1DB MEDIAN 2.17 NONE none  401 fwhm_med this | last_yr | all Mean fwhm, BLUE 437.0nm 1x1 0.4arcs slit
6 QC1DB MEDIAN 3.77 NONE none  101 fwhm_med this | last_yr | all Mean fwhm, BLUE 437.0nm 1x1 1.0arcs slit
7 +QC1DB none   NONE none  104 resol_med this | last_yr | all Correlation Resolution vs. Slit Width, BLUE 346.0nm 1x1, all slit widths (in black: theoretical values from User Manual)
8 +QC1DB none   NONE none  772 resol_med this | last_yr | all Correlation Resolution vs. Slit Width, BLUE 390.0nm 1x1, all slit widths (in black: theoretical values from User Manual)
9 +QC1DB none   NONE none  1039 resol_med this | last_yr | all Correlation Resolution vs. Slit Width, BLUE 437.0nm 1x1, all slit widths (in black: theoretical values from User Manual)
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:
(numbers below apply to this dataset)
resol_med
median:89.2NONE
thresholds:none 
N_data plotted:14
[click plot for closeup]
Plot 2
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:
(numbers below apply to this dataset)
resol_med
median:87.3NONE
thresholds:none 
N_data plotted:204
[click plot for closeup]
Plot 3
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:
(numbers below apply to this dataset)
resol_med
median:86.8NONE
thresholds:none 
N_data plotted:421
[click plot for closeup]
Plot 4
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:
(numbers below apply to this dataset)
fwhm_med
median:2.11NONE
thresholds:none 
N_data plotted:13
[click plot for closeup]
Plot 5
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:
(numbers below apply to this dataset)
fwhm_med
median:2.17NONE
thresholds:none 
N_data plotted:194
[click plot for closeup]
Plot 6
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:
(numbers below apply to this dataset)
fwhm_med
median:2.17NONE
thresholds:none 
N_data plotted:401
[click plot for closeup]
Plot 7
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:resol_med+
average:none
thresholds:none
N_data plotted:104
[click plot for closeup]
Plot 8
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:resol_med+
average:none
thresholds:none
N_data plotted:772
[click plot for closeup]
Plot 9
scores: not implemented
data source:uves_wave
(QC1 database)
dataset:resol_med+
average:none
thresholds:none
N_data plotted:1039
[click plot for closeup]
This plot

This is the trending report for monitoring UVES Resolution based on QC parameters derived from wavelength calibration frames (WAVE_ECH). It trends the measured resolution in the Health Check calibrations which are taken in one BLUE (437nm) and one RED (580nm) setting. It also displays a comparison of measured resolutions to the nominal values.

For these plots, no thresholds are set.

The individual plots describe:

  1. measured resolution for the BLUE 346.0nm, 1x1, 0.4 and 1.0 arcs slits
  2. measured resolution for the BLUE 390.0nm, 1x1, 0.4, 0.6 and 1.0 arcs slits
  3. measured resolution for the BLUE 437.0nm, 1x1, 0.4 arcs and 1.0 arcs slit
    For these plots the value of the lower threshold is set to the value given in the user manual. The Upper Chip red lower threshold is set to 95% because of charge diffusion.
  4. measured resolution as a function of slit-width for all BLUE 346.0nm, 1x1 data
  5. measured resolution as a function of slit-width for all BLUE 390.0nm, 1x1 data
  6. measured resolution as a function of slit-width for all BLUE 437.0nm, 1x1 data
In black dotted curve plots 4/5/6: calculated values from the User Manual (see table below). Find more information about UVES wavelength solution monitoring here.

Note: The Upper chip R is expected to be a bit lower due to charge diffusion. This was especially true of the old MIT/LL chip, seems to be better for the new Upper chip. The resolution for REDU is 5% lower than for REDL to allow for the charge diffusion

SLITW R_blue R_red(L) old_R_red(U)
#2.1 20000 20000 19000
#2.0 22000 22000 20900
#1.50 31000 31000 29450
#1.40 32000 32000 30400
#1.30 34000 34000 32300
#1.20 37000 37000 35150
#1.10 39000 39000 37050
#1.00 41000 42000 39900
#0.90 44000 46000 43700
#0.80 50000 53000 50350
#0.70 54000 60000 57000
#0.60 59000 68000 64600
#0.50 65000 78000 74100
#0.40 71000 91000 86450
#0.30 78000 108000 102600
#0.20 82000 129000 122550  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up