Plot ? |
Symb ? |
Source * |
Average ? |
Thresholds ? |
N_ data |
QC1 parameter |
Data downloads |
Remarks |
method |
value |
unit |
method |
value |
1 |
| QC1DB |
MEDIAN |
1.15 |
ADU |
VAL | 1.05,1.16 |
88 |
sigma_master |
this |
last_yr |
all
|
RON in master BIAS frame, BLUE CCD, readout mode 225kHz,1x1,low br |
2 |
| QC1DB |
MEDIAN |
1.68 |
ADU |
VAL | 1.58,1.74 |
108 |
sigma_master |
this |
last_yr |
all
|
same, for REDL CCD |
3 |
| QC1DB |
MEDIAN |
1.45 |
ADU |
VAL | 1.3,1.44 |
108 |
sigma_master |
this |
last_yr |
all
|
same, for REDU CCD |
4 |
| QC1DB |
MEDIAN |
2.21 |
ADU |
VAL | 2.09,2.32 |
86 |
sigma_master |
this |
last_yr |
all
|
same as Fig. 1, for readout mode 50kHz,2x2,high |
5 |
| QC1DB |
MEDIAN |
3.04 |
ADU |
VAL | 2.77,3.07 |
98 |
sigma_master |
this |
last_yr |
all
|
same, for REDL CCD |
6 |
| QC1DB |
MEDIAN |
2.67 |
ADU |
VAL | 2.49,2.76 |
98 |
sigma_master |
this |
last_yr |
all
|
same, for REDU CCD |
|
*Data sources: QC1DB: QC1 database; LOCAL: local data source
|
Plot 1 scores: 8|
| data source: | uves_bias (QC1 database) |
dataset: | sigma_master | |
median: | 1.15 | ADU |
fixed thresholds: | 1.05...1.16 | ADU |
N_data plotted: | 88 |
[click plot for closeup] |
Plot 2 scores: 8|
| data source: | uves_bias (QC1 database) |
dataset: | sigma_master | |
median: | 1.68 | ADU |
fixed thresholds: | 1.58...1.74 | ADU |
N_data plotted: | 108 |
[click plot for closeup] |
Plot 3 scores: 8|
| data source: | uves_bias (QC1 database) |
dataset: | sigma_master | |
median: | 1.45 | ADU |
fixed thresholds: | 1.3...1.44 | ADU |
N_data plotted: | 108 |
[click plot for closeup] |
Plot 4 scores: 8|
| data source: | uves_bias (QC1 database) |
dataset: | sigma_master | |
median: | 2.21 | ADU |
fixed thresholds: | 2.09...2.32 | ADU |
N_data plotted: | 86 |
[click plot for closeup] |
Plot 5 scores: 8|
| data source: | uves_bias (QC1 database) |
dataset: | sigma_master | |
median: | 3.04 | ADU |
fixed thresholds: | 2.77...3.07 | ADU |
N_data plotted: | 98 |
[click plot for closeup] |
Plot 6 scores: 8|
| data source: | uves_bias (QC1 database) |
dataset: | sigma_master | |
median: | 2.67 | ADU |
fixed thresholds: | 2.49...2.76 | ADU |
N_data plotted: | 98 |
[click plot for closeup] |
This plot
This is the trending report for the QC parameters derived from UVES BIAS frames.
These plots describe:
- the read noise (RON) of the BLUE, RED LOWER and RED UPPER CCD master frame in the 1x1 binning, 225kHz, low gain readout mode;
- the same for the 2x2 binning, 50kHz, high gain readout mode
The thresholds are set as -+5% of nominal values.
These nominal values are (in ADU)
- BLUE 1x1 1.11
- REDL 1x1 1.57
- REDU 1x1 1.47
- BLUE 2x2 2.21
- REDL 2x2 2.93
- REDU 2x2 2.63
The e- to ADU conversion used are
- BLUE 1x1 1.84 ADU
- REDL 1x1 1.59 ADU
- REDU 1x1 1.46 ADU
- BLUE 2x2 0.54 ADU
- REDL 2x2 0.52 ADU
- REDU 2x2 0.52 ADU
Find more information about UVES BIAS frames here.
Other QC parameters related to detector properties (like linearity, gain) are trended here.
General information
Click on any of the plots to see a close-up version.
The latest date is indicated on top of the plot.
If configured,
- data points belonging to the latest date are specially marked
- statistical averages are indicated by a solid line, and thresholds by broken lines
- outliers are marked by a red asterisk. They are defined as data points outside the
threshold lines
- "aliens" (= data points outside the plot Y limits) are marked by a red arrow (↑ or ↓)
- you can download the data for each parameter set if the 'Data downloads' link shows up
|