mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
VISIR:
sensitivity
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
VISIR trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-11-23T02:01:49 (UT)
  | now: 2024-11-23T07:12:28 (UT) muc03 QC pipeline: visir-4.4.3 (installed 2023-04-27)
same group: N_band Q_band
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
VISIR news:  
Report news:
DATE*: [?] 2024-11-16
report | NLT
2024-11-17
report | NLT
2024-11-18
report | NLT
2024-11-19
report | NLT
2024-11-20
report | NLT
2024-11-21
report | NLT
2024-11-22
no data | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw no CAL raw files
P...
CAL product quality:
products products products products products products no CAL products
 
transfer   ngas [?]

scores&comments | FULL | history ... | plot tutorial ... | contact | HC_docu... | quality events...               [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 33.6 mJy/1hr none  40 sensit this | last_yr | all sensitivity with Q1 filter, pfov=0.0453
2 QC1DB MEDIAN 35.8 mJy/1hr none  23 sensit this | last_yr | all sensitivity with Q2 filter, pfov=0.0453
3 QC1DB MEDIAN 85.9 mJy/1hr none  5 sensit this | last_yr | all sensitivity with Q3 filter, pfov=0.0453
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: not implemented
data source:visir_zp_img
(QC1 database)
dataset:sensit
median:33.6mJy/1hr
thresholds:none 
N_data plotted:40
[click plot for closeup]
Plot 2
scores: not implemented
data source:visir_zp_img
(QC1 database)
dataset:sensit
median:35.8mJy/1hr
thresholds:none 
N_data plotted:23
[click plot for closeup]
Plot 3
scores: not implemented
data source:visir_zp_img
(QC1 database)
dataset:sensit
median:85.9mJy/1hr
thresholds:none 
N_data plotted:5
[click plot for closeup]
This plot

Measurements of the sensitivity parameter for the three Q band filters. Note that starting with P95 only small-field (SF) observations with 0.0453 arcsec/pixel are offered.

Sensitivity is determined by observing a standard star and estimating the limiting flux expected in 1 hour "on-source" integration of a point-source detected with S/N of 10. It is given in units of mJy at 10 sigma in 1 hr.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up