mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
KMOS:
detector: dark
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
KMOS trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-11-22T15:50:09 (UT)
  | now: 2024-11-23T07:17:42 (UT) muc01 QC pipeline: kmos-4.0.3 (installed 2022-03-28)
same group: darkmed darkron badpix
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
KMOS news:  
Report news:
DATE*: [?] 2024-11-16
no data | NLT
2024-11-17
no data | NLT
2024-11-18
no data | NLT
2024-11-19
report | NLT
2024-11-20
report | NLT
2024-11-21
report | NLT
2024-11-22
report | NLT
r...
Raw CAL displays:
no CAL raw files no CAL raw files no CAL raw files no CAL raw files raw raw raw
P...
CAL product quality:
no CAL products no CAL products no CAL products no CAL products products products products
 
transfer   ngas [?]

scores&comments | FULL | history ... | plot tutorial ... | contact | HC_docu... | quality events...                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 2.59 ADU VAL2,5 87 ron_median this | last_yr | all median dark RON, DIT=60, detector 1
2 QC1DB MEDIAN 2.62 ADU VAL2,5 87 ron_median this | last_yr | all median dark RON, DIT=60, detector 2
3 QC1DB MEDIAN 2.87 ADU VAL2,5 87 ron_median this | last_yr | all median dark RON, DIT=60, detector 3
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 3|
data source:kmos_dark
(QC1 database)
dataset:ron_median
median:2.59ADU
fixed thresholds:2...5ADU
N_data plotted:87
[click plot for closeup]
Plot 2
scores: 3|
data source:kmos_dark
(QC1 database)
dataset:ron_median
median:2.62ADU
fixed thresholds:2...5ADU
N_data plotted:87
[click plot for closeup]
Plot 3
scores: 3|
data source:kmos_dark
(QC1 database)
dataset:ron_median
median:2.87ADU
fixed thresholds:2...5ADU
N_data plotted:87
[click plot for closeup]
This plot

Median RON (read-out noise) as measured with dark exposures.

The RON of the KMOS detectors depends on the used DIT: longer integration times have lower RON. We display here the values measured from the most commonly taken dark exposures with 60s DITs vs. times and the correlation of RON with DIT.

During the first two commissionings, measured RON values were significantly smaller than those measured during the following commissionings (March 2013 and later) and during normal operations (starting October 2013). Therefore, two separate data sets can be seen in the correlation plots.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up