CRIRES trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2014-07-22T11:32:51 (UT)
  | now: 2024-11-21T22:49:04 (UT) muc01 QC pipeline: crire-2.3.2 (installed 2014-03-18)
same group: darkmed ron_raw
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
CRIRES news:
Report news:
DATE*: [?] 2024-11-14
report | NLT
2024-11-15
report | NLT
2024-11-16
report | NLT
2024-11-17
report | NLT
2024-11-18
report | NLT
2024-11-19
report | NLT
2024-11-20
report | NLT
2024-11-21
no data | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw raw no CAL raw files
P...
CAL product quality:
products products products products products products products no CAL products
  last OPSLOG data:
2014-07-16*

last files ...
[an error occurred while processing this directive]
monitor

scores&comments | FULL | history ... | plot tutorial ... | contact                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB yes (1) MEDIAN 3.48 ADU VAL3.3,3.65 86 ron23 this | last_yr | all RON, chip 1, DIT=1
2 QC1DB yes (1) MEDIAN 3.66 ADU VAL3.4,3.9 86 ron23 this | last_yr | all RON, chip 2, DIT=1
3 QC1DB yes (1) MEDIAN 3.66 ADU VAL3.3,4.0 86 ron23 this | last_yr | all RON, chip 3, DIT=1
4 QC1DB yes (1) MEDIAN 4.08 ADU VAL3.7,4.5 86 ron23 this | last_yr | all RON, chip 4, DIT=1
5 QC1DB yes (1) MEDIAN 1.76 ADU VAL1.5,2.0 90 ron23 this | last_yr | all RON, chip 1, DIT=300
6 QC1DB yes (1) MEDIAN 2.15 ADU VAL1.8,2.5 90 ron23 this | last_yr | all RON, chip 2, DIT=300
7 QC1DB yes (1) MEDIAN 2.25 ADU VAL1.9,2.6 90 ron23 this | last_yr | all RON, chip 3, DIT=300
8 QC1DB yes (1) MEDIAN 2.1 ADU VAL1.7,2.5 90 ron23 this | last_yr | all RON, chip 4, DIT=300
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: OPSLOG data are included
Plot 1
scores: 8|
data source:crires_dark
(QC1 database)
dataset:ron23
median:3.48ADU
fixed thresholds:3.3...3.65ADU
N_data plotted:86
[click on plot for closeup]
Plot 2
scores: 8|
data source:crires_dark
(QC1 database)
dataset:ron23
median:3.66ADU
fixed thresholds:3.4...3.9ADU
N_data plotted:86
[click on plot for closeup]
Plot 3
scores: 8|
data source:crires_dark
(QC1 database)
dataset:ron23
median:3.66ADU
fixed thresholds:3.3...4.0ADU
N_data plotted:86
[click on plot for closeup]
Plot 4
scores: 8|
data source:crires_dark
(QC1 database)
dataset:ron23
median:4.08ADU
fixed thresholds:3.7...4.5ADU
N_data plotted:86
[click on plot for closeup]
Plot 5
scores: 8|
data source:crires_dark
(QC1 database)
dataset:ron23
median:1.76ADU
fixed thresholds:1.5...2.0ADU
N_data plotted:90
[click on plot for closeup]
Plot 6
scores: 8|
data source:crires_dark
(QC1 database)
dataset:ron23
median:2.15ADU
fixed thresholds:1.8...2.5ADU
N_data plotted:90
[click on plot for closeup]
Plot 7
scores: 8|
data source:crires_dark
(QC1 database)
dataset:ron23
median:2.25ADU
fixed thresholds:1.9...2.6ADU
N_data plotted:90
[click on plot for closeup]
Plot 8
scores: 8|
data source:crires_dark
(QC1 database)
dataset:ron23
median:2.1ADU
fixed thresholds:1.7...2.5ADU
N_data plotted:90
[click on plot for closeup]
This plot

Trending plot for CRIRES RON as determined from DARK exposures.

The read-out noise is measured from the difference of the 3rd and 2nd dark image using the following procedure:
- generate 100 13x13 pixel windows
- compute the pixel standard deviation in each window
- the read-out noise is the median of these measurements multiplied with sqrt(NDIT/2).

The trending shows RON from DIT=1sec and DIT=300sec exposures for all four chips.  


General information

Click on any of the plots to see a close-up version.

If applicable, the latest values from Paranal ops logs are plotted as blue dots. More information about those values, including a download option, is available in the OPS column.

The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines