Calibration Plan
Standard Calibrations
The most commonly used observing configurations of each instrument are calibrated by means of the Observatory's Calibration Plan. Calibration Plans for each instrument and mode are described in detail in their corresponding User Manual, and their execution times are not charged to the Service Mode programmes.
Additional Calibrations
Please check carefully in the User Manual whether or not the instrument configurations that you plan to use are included in the instrument's Calibration Plan. If this is not the case, you must provide additional OBs for calibration appropriate for each mode not included in the Calibration Plan. Their execution time is then charged to your time allocation.
User-supplied OBs for calibration are only executed once. You should assume that your science OBs will be executed completely independently of each other, possibly on different nights, and take into account that no user-supplied OB (for a science or a calibration target) will be executed more than once. You should submit enough OBs to cover that situation. For example, if you need to observe and flux-calibrate six targets in a filter that is not supported by the calibration plan of the instrument, you need to provide six OBs for photometric calibration to allow for the case in which each target is observed on a different night. This must be done even if the same calibration star can be used for all the science targets.
Please indicate in the User Comments fields of science OBs when a user-provided calibration must be executed contiguously.
'Concatenation containers' should be used within p2 to link science and calibration observations that must be executed contiguously.
Please note that there are two separate kinds of calibrations in operational terms:
- Calibrations that observe a reference celestial source, like a spectrophotometric standard star or a nearly-featureless star for the measurement of telluric lines. The OBs prepared for this kind of calibration are generically considered as normal OBs (to be prepared within an "OB" in p2), and their construction is the same as for other science OBs. However, depending of the instrument they use either templates whose name contains the indicator calib, or flags in the templates that indicate that the observation is intended for calibration.
- Calibrations with no celestial source associated, like flat fields taken with a flat field lamp or dark exposures. OBs for this kind of calibration should be prepared within the "CB" (Calibration Block) in p2, as no acquisition template needs to be specified in this case. Calibrations to be executed in daytime are the most common examples in this category.
ESPRESSO Calibration Plan
The ESPRESSO calibration plan can be found in the ESPRESSO User Manual.
This calibration plan should be sufficient for the majority of use cases. Observers interested in highest S/N may request additional flat-field frames. The calibration plan provides flat-field frames with a SNR of about 1000.
Please see the instrument news page for the availabilty of the Laser Frequency Comb (LFC).
Attached Calibrations
In case that the highest possible accuracy, such as flatfielding accuracy, is required, the use of attached calibrations at nighttime is possible (see the ESPRESSO User Manual). However, the execution time will be charged to the corresponding program.
Night time calibrations
All possible types of nighttime calibrations (spectrophotometric standards, radial-velocity standards, telluric standards, etc.) must be specifically requested by the users. OBs for these calibrations must be supplied by the user. The time needed to execute these OBs will be charged to the user. Please see the ESPRESSO User Manual for guidance on which template should be used for each type of calibration.