mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
PIONIER:
Transfer Function AT
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
PIONIER trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-11-23T05:38:35 (UT)
  | now: 2024-11-23T08:16:41 (UT) muc05 QC pipeline: pioni-3.9.6 (installed 2023-06-22)
same group: TF_AT_GR_HS TF_AT_GR_MS TF_AT_FR CP_AT
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
PIONIER news: 2023-02-06 Internal efficiency, median darks are outside KPI which is related to a pump not working. As operations continue, the thresholds have been adapted to show calibratibility and keep track of sudden anomalies.
Report news:
DATE*: [?] 2024-11-16
report | NLT
2024-11-17
report | NLT
2024-11-18
report | NLT
2024-11-19
report | NLT
2024-11-20
report | NLT
2024-11-21
report | NLT
2024-11-22
report | NLT
r...
Raw CAL displays:
n/a n/a n/a n/a n/a n/a n/a
P...
CAL product quality:
products products products products products products products
 
transfer   ngas [?]

scores&comments | FULL | history ... | plot tutorial ... | contact | HC_docu... | quality events...                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB none   NONE VAL0,2 1780 tfvis12_avg this | last_yr | all TF^2 High sensitivity Baseline12
2 QC1DB none   NONE VAL0,2 1778 tfvis13_avg this | last_yr | all Baseline 13
3 QC1DB none   NONE VAL0,2 1776 tfvis14_avg this | last_yr | all Bas1eline 4
4 QC1DB none   NONE VAL0,2 1779 tfvis23_avg this | last_yr | all Bas2eline 3
5 QC1DB none   NONE VAL0,2 1777 tfvis24_avg this | last_yr | all Bas2eline 4
6 QC1DB none   NONE VAL0,2 1775 tfvis34_avg this | last_yr | all Bas3eline 4

*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: not implemented
data source:pionier_fringe
(QC1 database)
dataset:tfvis12_avg
average:none
fixed thresholds:0...2NONE
N_data plotted:1780
[click plot for closeup]
Plot 2
scores: not implemented
data source:pionier_fringe
(QC1 database)
dataset:tfvis13_avg
average:none
fixed thresholds:0...2NONE
N_data plotted:1778
[click plot for closeup]
Plot 3
scores: not implemented
data source:pionier_fringe
(QC1 database)
dataset:tfvis14_avg
average:none
fixed thresholds:0...2NONE
N_data plotted:1776
[click plot for closeup]
Plot 4
scores: not implemented
data source:pionier_fringe
(QC1 database)
dataset:tfvis23_avg
average:none
fixed thresholds:0...2NONE
N_data plotted:1779
[click plot for closeup]
Plot 5
scores: not implemented
data source:pionier_fringe
(QC1 database)
dataset:tfvis24_avg
average:none
fixed thresholds:0...2NONE
N_data plotted:1777
[click plot for closeup]
Plot 6
scores: not implemented
data source:pionier_fringe
(QC1 database)
dataset:tfvis34_avg
average:none
fixed thresholds:0...2NONE
N_data plotted:1775
[click plot for closeup]
This plot

This is the trending plot using PIONIER QC parameters derived from CALIB_FRINGE frames.

The pipeline calculates the TF using the diameter of the object extracted from the faint calibrator catalog (if found)
. Each of the plot represent the TF^2 measured on one baseline.
For the ATs, the scatter of the measured TF^2 could come from the atmosphere or from the fact that the telescope positioned on a given station (1,2,3 or 4) could be different (AT1,2,3 or 4) and the baseline length can also be very different
. An outlier could come from a wrong diameter in the calibrator catalog.
This information is given in the "data Download" link.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up