Calibration data for FORS1 are taken for two purposes:
assess the instrument status
reduce science data
Generally, FORS1 calibration data are taken as daytime calibrations.
On user request, also nighttime calibrations can be
obtained as part of the Observing Block which generates the
SCIENCE data ("attached calibrations").
Please, refer to Chapter 4 of the FORS User Manual for details on the current calibration plan.
The calibration data taken obtained in the supported modes are processed by QC
Garching to obtain calibration solutions. These solutions
come as:
All 4-port frames belonging to pipeline-supported modes are processed
using FTU (Fits Translation
Utility) and a set of IRAF/IDL readable keywords is added to the the FITS
headers. In particular a full description of the port structure, including
pre-scan regions, is included.
IMPORTANT NOTE: all FORS1 pipeline products do not have the pre-scan regions which are
chopped-off during the reduction process. As a consequence, master calibrations/reduced science
files start at X pixel coordinates @17 and stop at @2064. In order to use the pipeline
master calibrations with raw frames, one needs to perform the trimming
of input images in the same way.
RAW CALIBRATION DATA
The following types of raw calibration data exist for the supported modes:
IMAGING
Frame Type
DPR Type
DPR Tech
Purpose
BIAS
BIAS
IMAGE
bias level, RON
DARK
DARK
IMAGE
dark current
SKY_FLAT_IMG
FLAT,SKY
IMAGE
flat field of science data
SCREEN_FLAT_IMG
FLAT,LAMP
IMAGE
gain, lamp efficiency, fixed pattern noise
STANDARD_IMG
STD
IMAGE
photometric
zeropoints, extinction and
colour coefficients
LSS
Frame Type
DPR Type
DPR Tech
Purpose
BIAS
BIAS
IMAGE
bias level, RON
SCREEN_FLAT_LSS/LAMP_LSS
FLAT,LAMP/WAVE,LAMP
SPECTRUM
flat field of science data, slit transfer function, 2D dispersion solution and distortion correction
STANDARD_LSS
STD
SPECTRUM
spectrophotometric calibration
MOS
Frame Type
DPR Type
DPR Tech
Purpose
BIAS
BIAS
IMAGE
bias level, RON
SCREEN_FLAT_MOS/LAMP_MOS
FLAT,LAMP/WAVE,LAMP
MOS
flat field of science data, slit transfer function, 2D dispersion solution and distortion correction
STANDARD_MOS
STD
MOS
spectrophotometric calibration
MASTER CALIBRATION DATA AND PRODUCTS
The following types of master calibration data are produced for the
supported modes:
reduce data and apply wavelength correction, correct sky background and extract spectrum
1
extracted spectrum with errors, sky background
CALIBRATION CASCADE
Like for all astronomical data, the reduction of FORS1 calibration frames requires a cascading scheme where the
mutual dependencies of products and raw frames have to be respected. Master Biases can be produced from raw
files only, but all the other products require the prior generation of other calibration products. The proper sequence of
all these production steps is called the calibration cascade.
NAMING SCHEME
After production and certification, the calibration products are renamed after a scheme which includes type, creation date and relevant FORS1 settings.