VISIR trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-11-24T02:01:35 (UT)
  | now: 2024-11-24T08:16:04 (UT) muc03 QC pipeline: visir-4.4.3 (installed 2023-04-27)
same group: read-noise dark det-temp
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
VISIR news:  
Report news:
DATE*: [?] 2024-11-17
report | NLT
2024-11-18
report | NLT
2024-11-19
report | NLT
2024-11-20
report | NLT
2024-11-21
report | NLT
2024-11-22
report | NLT
2024-11-23
no data | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw no CAL raw files
P...
CAL product quality:
products products products products products products no CAL products
 
transfer   ngas [?]

scores&comments | history ... | plot tutorial ... | contact | HC_docu... | quality events...                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB FIXED 9.1 K VAL8.9,9.3 90 ins_sens30_val this | last_yr | all temperature of the imaging detector
2 QC1DB FIXED 9.1 K VAL8.9,9.3 89 ins_sens31_val this | last_yr | all temperature of the spectroscopic detector
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 9|
data source:visir_dark_img
(QC1 database)
dataset:ins_sens30_val
fixed average:9.1K
fixed thresholds:8.9...9.3K
N_data plotted:90
[click plot for closeup]
Plot 2
scores: 8|
data source:visir_dark_spc
(QC1 database)
dataset:ins_sens31_val
fixed average:9.1K
fixed thresholds:8.9...9.3K
N_data plotted:89
[click plot for closeup]
This plot

This is a trending plot monitoring temperatures of the VISIR imaging and spectroscopic detectors. As of May 12, 2010 the temperature of the imaging detector is read from keyword INS_TEMP30_VAL and of the spectroscopic detector from keyword INS_TEMP31_VAL. Before that date different keywords were used - INS_SENS30_VAL and INS_SENS31_VAL respectively.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up