VIRCAM trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2023-06-12T10:25:50 (UT)
  | now: 2024-12-22T18:24:10 (UT) muc07 QC pipeline: vcam-2.3.9 (installed 2022-03-17)
same group: RESETMED RESETRMS RESETDIFFRMS   temperatures
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
VIRCAM news: On March 6th, 2023, VIRCAM has been decommissioned after 13 years of operations.
Report news:
DATE*: [?] 2023-06-05
no data | NLT
2023-06-06
no data | NLT
2023-06-07
no data | NLT
2023-06-08
no data | NLT
2023-06-09
no data | NLT
2023-06-10
no data | NLT
2023-06-11
no data | NLT
r...
Raw CAL displays:
no CAL raw files no CAL raw files no CAL raw files no CAL raw files no CAL raw files no CAL raw files no CAL raw files
P...
CAL product quality:
no CAL products no CAL products no CAL products no CAL products no CAL products no CAL products no CAL products
 
transfer   ngas [?]

scores&comments | history ... | plot tutorial ... | contact | HC_docu... | quality events...                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 502 ADU VAL450,525 31 qc_resetrms this | last_yr | all RESETRMS, average over all 16 detectors
2 QC1DB MEDIAN 458 ADU VAL250,1100 496 qc_resetrms this | last_yr | all RESETRMS, values of all detetcors
2 QC1DB none   ADU none  31 qc_resetrms this | last_yr | all RESETRMS, values of detector 2
2 QC1DB none   ADU none  31 qc_resetrms this | last_yr | all RESETRMS, values of detector 3
2 QC1DB none   ADU none  31 qc_resetrms this | last_yr | all RESETRMS, values of detector 13
3 QC1DB MEDIAN 148 ADU VAL120,190 31 qc_resetrms this | last_yr | all RESETRMS, stdev over all detectors
4 LOCAL none   ADU none  488 qc_resetrms n/a Score per Detector: Green = within the thresholds
4 LOCAL none   ADU none  8 qc_resetrms n/a Score per Detector: Red = beyond thresholds
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 3|
data source:vircam_reset
(QC1 database)
dataset:qc_resetrms
median:502ADU
fixed thresholds:450...525ADU
N_data plotted:31
[click plot for closeup]
Plot 2
scores: not implemented
data source:vircam_reset
(QC1 database)
dataset:
(numbers below apply to this dataset)
qc_resetrms
median:458ADU
fixed thresholds:250...1100ADU
N_data plotted:496
[click plot for closeup]
Plot 3
scores: 3|
data source:vircam_reset
(QC1 database)
dataset:qc_resetrms
median:148ADU
fixed thresholds:120...190ADU
N_data plotted:31
[click plot for closeup]
Plot 4
scores: not implemented
data source:local data source
dataset:
(numbers below apply to this dataset)
qc_resetrms
average:none
thresholds:none
N_data plotted:488
[click plot for closeup]
This plot

Trending plot for RMS noise in combined reset frame.

QC parameters is derived from each detector of daily RESET frames.

Name:
Fits header: QC.RESETRMS,
QC1 DB: qc_resetrms

Description:
RMS noise of MASTER_RESET pipeline product frame

Setup:
DET.DIT = MINDIT

These plots describe:
Fig. 1. Average view: The QC parameter averaged over all detectors. This view is sensitive to common changes.
Fig. 2. Resolved View: The QC parameter resolved for all detectors.
Fig. 3. Scatter View: The standard deviation of the QC parameter, obtained from the sample of all detectors. This view is sensitive changes of individual detectors.
Fig. 4. Scoring View: Scoring values of all 16 QC parameters. Green = within the acceptance thresholds. Red = beyond the acceptance thresholds.

In plot (2) chip #13 shows the largest RMS among all chips.

Find more information on the VIRCAM RESET tutorial page.

 


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up