Plot ? |
Symb ? |
Source * |
Average ? |
Thresholds ? |
N_ data |
QC1 parameter |
Data downloads |
Remarks |
method |
value |
unit |
method |
value |
1 |
| QC1DB |
MEDIAN |
11.7 |
ADU |
VAL | 10.2,13.2 |
13 |
current_dark |
this |
last_yr |
all
|
dark current in Parasitic Light, readout mode 50kHz,2x2,high; BLUE CCD |
3 |
| QC1DB |
MEDIAN |
3.89 |
ADU |
VAL | 2.2,5.8 |
14 |
current_dark |
this |
last_yr |
all
|
same, for REDL CCD |
4 |
| QC1DB |
MEDIAN |
2.73 |
ADU |
VAL | 0.1,4.9 |
14 |
current_dark |
this |
last_yr |
all
|
same, for REDU CCD |
|
*Data sources: QC1DB: QC1 database; LOCAL: local data source
|
Plot 1 scores: not implemented
| data source: | uves_dark (QC1 database) |
dataset: | current_dark | |
median: | 11.7 | ADU |
fixed thresholds: | 10.2...13.2 | ADU |
N_data plotted: | 13 |
[click plot for closeup] |
Plot 3 scores: not implemented
| data source: | uves_dark (QC1 database) |
dataset: | current_dark | |
median: | 3.89 | ADU |
fixed thresholds: | 2.2...5.8 | ADU |
N_data plotted: | 14 |
[click plot for closeup] |
Plot 4 scores: not implemented
| data source: | uves_dark (QC1 database) |
dataset: | current_dark | |
median: | 2.73 | ADU |
fixed thresholds: | 0.1...4.9 | ADU |
N_data plotted: | 14 |
[click plot for closeup] |
This plot
This is the trending report for UVES QC parameters derived from Parasitic
Light frames.
These frames are taken approximately once per month, as a single exposure of
3600 secs in a dark enclosure BUT with the CCD shutter OPEN (as compared to
the DARK frames where the CCD shutter is closed). These data are de-biased
and combined into a MASTER_DARK, in which the average dark level is measured
(in ADU/px/hr). [There are also shorter exposures, they are subject to false
results. If you find an outlier, check first the exposure time under 'Data
downloads'!]
The purpose of these exposures is to detect low-level contaminating light
sources (e.g. LEDs) originating INSIDE the instrument.
The thresholds are set to the values defined by the KPI document from IS
These plots describe:
- the DARK current in master PARASITIC_DARK (or PDARK) frames of the BLUE|REDL|REDU chips in the 1x1 binning, 225kHz, low gain readout mode
- same for the BLUE|REDL|REDU chips in the 2x2 binning, 50kHz, high gain readout mode.
General information
Click on any of the plots to see a close-up version.
The latest date is indicated on top of the plot.
If configured,
- data points belonging to the latest date are specially marked
- statistical averages are indicated by a solid line, and thresholds by broken lines
- outliers are marked by a red asterisk. They are defined as data points outside the
threshold lines
- "aliens" (= data points outside the plot Y limits) are marked by a red arrow (↑ or ↓)
- you can download the data for each parameter set if the 'Data downloads' link shows up
|