PIONIER news:
2023-02-06 Internal efficiency, median darks are outside KPI which is related to a pump not working. As operations continue, the thresholds have been adapted to show calibratibility and keep track of sudden anomalies.
The pipeline calculates the central wavelength as well as the average value of the effective bandwidth for the 2 dispersive modes : GRISM and FREE.
Plots 3 and 6 (peak_to_peak) monitor the maximum variation of the effective band between the different baselines
The detector polarization voltages (DET.POLAR) are :
High Sensitivity : 7100 mV
Medium_ Sensitivity: 5000 mV
Low_Sensitivity : 1000 mV
General information
Click on any of the plots to see a close-up version.
The latest date is indicated on top of the plot.
If configured,
data points belonging to the latest date are specially marked
statistical averages are indicated by a solid line, and thresholds by broken lines
outliers are marked by a red asterisk. They are defined as data points outside the
threshold lines
"aliens" (= data points outside the plot Y limits) are marked by a red arrow (↑ or ↓)
you can download the data for each parameter set if the 'Data downloads' link shows up