NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
PIONIER news:
2023-02-06 Internal efficiency, median darks are outside KPI which is related to a pump not working. As operations continue, the thresholds have been adapted to show calibratibility and keep track of sudden anomalies.
The gain is monitored for the full detector (320x255) and for the
"science" pixels (26 windows of 6x1 pixels).
Only the gain across the full detector is scored
The gain has been measured after installation of RAPID (december 2014):
Low_Sensitivity:
- det_polar=1000mv gain=1.398 e-/ADU
- det_polar=3000mv gain=0.861 e-/ADU
Medium_Sensitivity:
- det_polar=5000mv gain=0.296 e-/ADU
High Sensitivity:
- det_polar=7100mv gain=0.107 e-/ADU
General information
Click on any of the plots to see a close-up version.
The latest date is indicated on top of the plot.
If configured,
data points belonging to the latest date are specially marked
statistical averages are indicated by a solid line, and thresholds by broken lines
outliers are marked by a red asterisk. They are defined as data points outside the
threshold lines
"aliens" (= data points outside the plot Y limits) are marked by a red arrow (↑ or ↓)
you can download the data for each parameter set if the 'Data downloads' link shows up