[ ESO ]

UVES Scientific CCD System

HOME INDEX SEARCH HELP NEWS

Calibrated Readout Modes Available at the Telescope

UVES Blue Arm CCD        (see also the   Data for UVES Blue   before   the upgrade on 13/Oct/2004))

Device :       EEV CCD-44-82 "PAVAROTTI"
CCD carrier temperature : 153 K   (accoustic alarms set at 143 K and 163 K)
CCD operating temperature : - 115 °C   (est.)
CFC cold plate temperature : - 170 °C

Read out units 225_lg_1x1_1pt 225_lg_1x2_1pt 50_hg_2x2_1pt 50_hg_2x3_1pt
Port used - left left left left
Conversion factor e- / ADU 1.84 1.74 0.54 0.54
Readout noise e- 4.03 3.81 2.36 2.31
Dark current
(system up and dark > 24 h)
e- / pix / h < 0.2 - < 0.36 -
Bias level ADU 201 201 150 149
Linearity % 0.77 0.63 0.47 0.32
Cosmic ray hit rate
(events > 30 e-)
hits / cm2 / h 26 - 22 -





UVES Red Arm CCD

Device :       EEV CCD-44-82 "STING"
CCD carrier temperature : 135 K   (accoustic alarms set at 133 and 140 K)
CCD operating temperature : - 115 °C   (est.)
Cold plate temperature : - 185 °C

Read out units 225_lg_1x1_1pt 225_lg_1x2_1pt 50_hg_2x2_1pt 50_hg_2x3_1pt
Port used - left left left left
Conversion factor e- / ADU 1.59 1.59 0.52 0.52
Readout noise e- 3.42 3.42 1.92 1.92
Dark current
(system up and dark > 24 h)
e- / pix / h - - 0.4 0.4
Bias level ADU 199 199 201 201
Linearity % + 0.42 / - 0.48 + 0.42 / - 0.48 + 0.15 / - 0.12 + 0.15 / - 0.12
Cosmic ray hit rate
(events > 30 e-)
hits / cm2 / h 112 112 113 113
CTE (X) - 0.999998 @ 6 x 104 e- 0.999998 @ 6 x 104 e- 0.999890 @ 2 x 102 e-
0.999997 @ 3 x 104 e-
0.999890 @ 2 x 102 e-
0.999997 @ 3 x 104 e-



Device :       MIT CCID-20 "NIGEL"    (Serial No. W7C2)
CCD carrier temperature : 135 K
CCD operating temperature : - 118 °C   (est.)
Cold plate temperature : - 185 °C

Read out units 225_lg_1x1_1pt 225_lg_1x2_1pt 50_hg_2x2_1pt 50_hg_2x3_1pt
Port used - right right right right
Conversion factor e- / ADU 1.48 1.48 0.57 0.57
Readout noise e- 3.69 3.69 2.87 2.87
Dark current
(system up and dark > 24 h)
e- / pix / h - - 0.6 0.6
Bias level ADU 191 191 181 - 186 181 - 186
Linearity % + 0.10 / - 0.10 + 0.10 / - 0.10 + 0.03 / - 0.03 + 0.03 / - 0.03
Cosmic ray hit rate
(events > 30 e-)
hits / cm2 / h 108 108 102 102
CTE (X) - > 0.999998 @ 8 x 104 e- > 0.999998 @ 8 x 104 e- - -


Back to   "UVES Scientific CCD System"

Send comments to   <odt@eso.org>
Last update: Jan 27, 2010
Webdesign by   EVI