UVES Scientific CCD System |
|||||
|
Device : | EEV CCD-44-82 "PAVAROTTI" | |
CCD carrier temperature : | 153 K (accoustic alarms set at 143 K and 163 K) | |
CCD operating temperature : | - 115 °C (est.) | |
CFC cold plate temperature : | - 170 °C |
Read out | units | 225_lg_1x1_1pt | 225_lg_1x2_1pt | 50_hg_2x2_1pt | 50_hg_2x3_1pt |
Port used | - | left | left | left | left |
Conversion factor | e- / ADU | 1.84 | 1.74 | 0.54 | 0.54 |
Readout noise | e- | 4.03 | 3.81 | 2.36 | 2.31 |
Dark current (system up and dark > 24 h) |
e- / pix / h | < 0.2 | - | < 0.36 | - |
Bias level | ADU | 201 | 201 | 150 | 149 |
Linearity | % | 0.77 | 0.63 | 0.47 | 0.32 |
Cosmic ray hit rate (events > 30 e-) |
hits / cm2 / h | 26 | - | 22 | - |
Device : | EEV CCD-44-82 "STING" | |
CCD carrier temperature : | 135 K (accoustic alarms set at 133 and 140 K) | |
CCD operating temperature : | - 115 °C (est.) | |
Cold plate temperature : | - 185 °C |
Read out | units | 225_lg_1x1_1pt | 225_lg_1x2_1pt | 50_hg_2x2_1pt | 50_hg_2x3_1pt |
Port used | - | left | left | left | left |
Conversion factor | e- / ADU | 1.59 | 1.59 | 0.52 | 0.52 |
Readout noise | e- | 3.42 | 3.42 | 1.92 | 1.92 |
Dark current (system up and dark > 24 h) |
e- / pix / h | - | - | 0.4 | 0.4 |
Bias level | ADU | 199 | 199 | 201 | 201 |
Linearity | % | + 0.42 / - 0.48 | + 0.42 / - 0.48 | + 0.15 / - 0.12 | + 0.15 / - 0.12 |
Cosmic ray hit rate (events > 30 e-) |
hits / cm2 / h | 112 | 112 | 113 | 113 |
CTE (X) | - | 0.999998 @ 6 x 104 e- | 0.999998 @ 6 x 104 e- | 0.999890 @ 2 x 102 e- 0.999997 @ 3 x 104 e- |
0.999890 @ 2 x 102 e- 0.999997 @ 3 x 104 e- |
Device : | MIT CCID-20 "NIGEL" (Serial No. W7C2) | |
CCD carrier temperature : | 135 K | |
CCD operating temperature : | - 118 °C (est.) | |
Cold plate temperature : | - 185 °C |
Read out | units | 225_lg_1x1_1pt | 225_lg_1x2_1pt | 50_hg_2x2_1pt | 50_hg_2x3_1pt |
Port used | - | right | right | right | right |
Conversion factor | e- / ADU | 1.48 | 1.48 | 0.57 | 0.57 |
Readout noise | e- | 3.69 | 3.69 | 2.87 | 2.87 |
Dark current (system up and dark > 24 h) |
e- / pix / h | - | - | 0.6 | 0.6 |
Bias level | ADU | 191 | 191 | 181 - 186 | 181 - 186 |
Linearity | % | + 0.10 / - 0.10 | + 0.10 / - 0.10 | + 0.03 / - 0.03 | + 0.03 / - 0.03 |
Cosmic ray hit rate (events > 30 e-) |
hits / cm2 / h | 108 | 108 | 102 | 102 |
CTE (X) | - | > 0.999998 @ 8 x 104 e- | > 0.999998 @ 8 x 104 e- | - | - |