mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
XSHOOTER:
detector: parameters
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
XSHOOTER trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-11-22T22:35:28 (UT)
  | now: 2024-11-23T07:08:01 (UT) muc02 QC pipeline: xshoo-3.6.5a (installed 2023-09-18)
same group: gain linearity max_flux dark
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
XSHOOTER news:  
Report news:
DATE*: [?] 2024-11-16
report | NLT
2024-11-17
report | NLT
2024-11-18
report | NLT
2024-11-19
report | NLT
2024-11-20
report | NLT
2024-11-21
report | NLT
2024-11-22
no data | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw no CAL raw files
P...
CAL product quality:
products products products products products products no CAL products
 
transfer   ngas [?]

scores&comments | FULL | history ... | plot tutorial ... | contact | HC_docu... | quality events...               [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB none   ADU/sec VAL-0.0005,0.001 17 median_dark this | last_yr | all 100k/1pt/hg
2 QC1DB none   ADU/sec none  17 sigma_dark this | last_yr | all 100k/1pt/hg
3 QC1DB none   ADU/sec VAL-0.0005,0.001 14 median_dark this | last_yr | all 400k/1pt/lg
4 QC1DB none   ADU/sec none  14 sigma_dark this | last_yr | all 400k/1pt/lg
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 1|
data source:xshooter_dark
(QC1 database)
dataset:median_dark
average:none
fixed thresholds:-0.0005...0.001ADU/sec
N_data plotted:17
[click plot for closeup]
Plot 2
scores: not implemented
data source:xshooter_dark
(QC1 database)
dataset:sigma_dark
average:none
thresholds:none
N_data plotted:17
[click plot for closeup]
Plot 3
scores: 1|
data source:xshooter_dark
(QC1 database)
dataset:median_dark
average:none
fixed thresholds:-0.0005...0.001ADU/sec
N_data plotted:14
[click plot for closeup]
Plot 4
scores: not implemented
data source:xshooter_dark
(QC1 database)
dataset:sigma_dark
average:none
thresholds:none
N_data plotted:14
[click plot for closeup]
This plot

Dark frames are taken once per month with 1x1 bin for both low and high gain. The typical exposure time is 3600 sec. The XSHOOTER pipeline generates master DARKs from input raw DARKs with master BIAS subtracted. The dark current is computed from the median master dark frame level in ADU. For more information see the plot tutorial page.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up