Dark frames are typically measured once a month in two read-out modes:

  • low gain, 1x1 binning, 2148x2440, used for imaging,
  • high gain, 1x1 binning, 2148x4096, used for spectroscopy.

All QC parameters are computed in the central 1600x1800 pixels of each CCD.

top Dark current

QC1 parameters

parameter QC1 database: table, name procedure
dark current vimos_bias, dark_avg computed in ADU/hr/pix from the average level of the master and then transformed into e/hr/pix using the conversion factor CONAD

Trending

The dark current is trended for each quadrant and for each read-out mode separately.

History

October 2010: new detectors with lower dark current


 