SPHERE trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-12-22T16:11:00 (UT)
  | now: 2024-12-23T02:46:03 (UT) muc03 QC pipeline: spher-0.56.0 (installed 2024-09-26)
same group: counts_YJ counts_YJH NumBadPix_YJ NumBadPix_YJH
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
SPHERE news: Note that the current Health Check monitor runs with an incomplete pipeline and an incomplete set of QC1 parameters. QCG can currently not guarantee completeness and significance at the usual level.
Report news:
DATE*: [?] 2024-12-16
report | NLT
2024-12-17
report | NLT
2024-12-18
report | NLT
2024-12-19
report | NLT
2024-12-20
report | NLT
2024-12-21
report | NLT
2024-12-22
report | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw no CAL raw files
P...
CAL product quality:
products products products products products products no CAL products
 
transfer   ngas [?]

scores&comments | FULL | history ... | plot tutorial ... | contact | HC_docu... | quality events...               [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB none   ADU VAL8000,10500 29 qc_lam_flux_avg_1 this | last_yr | all narrow band laser lamp 1 counts in central layer of second raw frame cube
2 QC1DB none   ADU VAL21500,23000 29 qc_lam_flux_avg_1 this | last_yr | all narrow band laser lamp 2 counts in central layer of second raw frame cube
3 QC1DB none   ADU VAL11000,13000 29 qc_lam_flux_avg_1 this | last_yr | all broad band halogen lamp counts in central layer of second raw frame cube
4 QC1DB none   ADU VAL8500,10500 29 qc_lam_flux_avg_1 this | last_yr | all narrow band laser lamp 3 counts in central layer of second raw frame cube
5 QC1DB none   ADU VAL7000,8000 29 qc_lam_flux_avg_1 this | last_yr | all narrow band laser lamp 4 counts in central layer of second raw frame cube
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 2|
data source:sphere_ifs_dflat
(QC1 database)
dataset:qc_lam_flux_avg_1
average:none
fixed thresholds:8000...10500ADU
N_data plotted:29
[click plot for closeup]
Plot 2
scores: 2|
data source:sphere_ifs_dflat
(QC1 database)
dataset:qc_lam_flux_avg_1
average:none
fixed thresholds:21500...23000ADU
N_data plotted:29
[click plot for closeup]
Plot 3
scores: 2|
data source:sphere_ifs_dflat
(QC1 database)
dataset:qc_lam_flux_avg_1
average:none
fixed thresholds:11000...13000ADU
N_data plotted:29
[click plot for closeup]
Plot 4
scores: 2|
data source:sphere_ifs_dflat
(QC1 database)
dataset:qc_lam_flux_avg_1
average:none
fixed thresholds:8500...10500ADU
N_data plotted:29
[click plot for closeup]
Plot 5
scores: 2|
data source:sphere_ifs_dflat
(QC1 database)
dataset:qc_lam_flux_avg_1
average:none
fixed thresholds:7000...8000ADU
N_data plotted:29
[click plot for closeup]
This plot

Imaging flats of the IFS arm.

The template generates two raw frames, each one is a fits cube ( NAXIS = 3 ) with five layers ( NAXIS3 = 5 ) corresponding to NDIT=5 and with QC parameters:
QC.LAMP.FLUX.AVG.0 to QC.LAMP.FLUX.AVG.4 (from the first raw frame cube) and QC parameters:
QC.LAMP.FLUX.AVG.5 to QC.LAMP.FLUX.AVG.9 (from the second raw frame cube).

Here we plot qc_lam_flux_avg_1 (= QC.LAMP.FLUX.AVG.7 = middle layer of second cube).

LAMP_1 ... LAMP_4 are monochromatic LASER. LAMP_BB is a broad band halogen lamp.

Currently data from two different templates are shown with slightly different setups, the 'Calibration Flats for Science' template and the 'Static Weekly Detector full calibration' template.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up