mirror sites: PL (internal link) HQ     [?]

Quality Control and
Data Processing


CAL | HC | refs | QC
QUALITY CONTROL
    HOME
HAWK-I QC
Trending & QC1
   dark
Pipeline
HAWKI ...
QC links:
HAWKI: Dark
DARK median | DARK ron | DARK reset anomaly | DARK ron vs. DIT

 
HC PLOTS
Dark median
Dark ron
Reset anomaly
Ron vs. DIT
QC1 database (advanced users): browse | plot

The HAWKI detectors' dark trending is based on measurement of the QC1 parameters from the daily DARK calibration data. They are obtained with dedicated Health Check OBs. Four sets of HAWK-I DARK frames with four different DIT/NDIT/NEXP combinations are taken daily for monitoring purposes. These data can of course also be used for calibration of scientific data if necessary. The four DIT/NDIT/NEXP combinations are: 1.6762/1/10, 2.0/1/10, 60.0/1/5 and 300.0/1/3. The Median levels and ReadOut Noises trends are monitored for 2.0 and 300 second DITs and the Reset Anomolies and the ReadOut Noise vs. DIT correllation relations are monitored from the minimum DIT (currently 1.6762sec) up to 300sec.

Master DARK: This is an example of a typical master dark frame. It contains 4 extensions for each of the 4 chips of HAWK-I. Here shown together.


DARK median
DARK median | DARK ron | DARK reset anomaly | DARK ron vs. DIT

QC1_parameters

FITS key QC1 database: table, name definition class* HC_plot** more docu
QC.DARK.MED hawki_dark..qc_dark_med median value across master dark HC [docuSys coming]
*Class: KPI - instrument performance; HC - instrument health; CAL - calibration quality; ENG - engineering parameter
**There might be more than one.

Trending

The dark level of the master dark in each of the 4 detectors is monitored.

Scoring&thresholds DARK median

The dark_median in each detector is scored individually, with dynamic thresholds following the median average of the current period. The only requirement on the dark current is stability on the calibration timescale.

History

Date event
2014-02-01 Bright darks on Feb 01 due to high detector temperature. Instrumentation worked on it and problem was fixed.

Algorithm DARK median

A list of dark frames (at least 5 input frames, usually 10) is combined into a mean dark product. The mean value is estimated across the whole detector for each chip separately.


DARK ron
DARK median | DARK ron | DARK reset anomaly | DARK ron vs. DIT

QC1_parameters

FITS key QC1 database: table, name definition class* HC_plot** more docu
QC.RON2 hawki_dark..qc_ron23 ReadOut Noise from (second-third) RAW frames HC [docuSys coming]
*Class: KPI - instrument performance; HC - instrument health; CAL - calibration quality; ENG - engineering parameter
**There might be more than one.

Trending

The read-out-noise estimated from the pair of the 2nd and 3rd exposure is monitored for each of the 4 detectors. Only the 2.0 and 300 second DIT calibration frames are used.

Scoring&thresholds DARK ron

The read-out-noise in each detector is scored individually, with dynamic thresholds following the median average of the current period. The only requirement on the dark current is stability on the calibration timescale.

History

Date event
2016-11-19 new, re-designed CASU pipeline was adopted in the QC system

Algorithm DARK ron

A list of dark frames (at least 5 input frames, usually 10) is combined into a mean dark product. Read-out-noise is estimated from each pair of the input frames.


DARK reset anomaly
DARK median | DARK ron | DARK reset anomaly | DARK ron vs. DIT

The Reset Anomoly is a feature of certain detectors whereby after the detector reset there is a period of time during which the rate of dark current accumulation varies before settling into a (more or less) constant rate of accumulation, e.g. as seen in ISAAC.

Based on long term trending it looks like HAWK-I does NOT suffer from a Reset Anomoly.

QC1_parameters

FITS key QC1 database: table, name definition class* HC_plot** more docu
QC.DARK.MED hawki_dark..qc_dark_med median value across master dark HC [docuSys coming]
*Class: KPI - instrument performance; HC - instrument health; CAL - calibration quality; ENG - engineering parameter
**There might be more than one.

Trending

The dark median values are trended as a function of data integration time (DIT).

Scoring&thresholds DARK reset anomaly

No scoring

History

Date event
2016-11-19 new, re-designed CASU pipeline was adopted in the QC system

Algorithm DARK reset anomaly

No algorithm applied


DARK ron vs. DIT
DARK median | DARK ron | DARK reset anomaly | DARK ron vs. DIT

QC1_parameters

FITS key QC1 database: table, name definition class* HC_plot** more docu
QC.RON2 hawki_dark..qc_ron23 ReadOut Noise from (second-third) RAW frames HC [docuSys coming]
*Class: KPI - instrument performance; HC - instrument health; CAL - calibration quality; ENG - engineering parameter
**There might be more than one.

Trending

This correlation plot monitors for any correllation between read-out-noise (RON) and data integration time (DIT).

Scoring&thresholds DARK ron vs. DIT

No scoring

History

Date event
2016-11-19 new, re-designed CASU pipeline was adopted in the QC system

Algorithm DARK ron vs. DIT

No algorithm applied


Send comments to <qc_hawki@eso.org> powered by QC [webCMS v1.0]