mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
GRAVITY:
detector
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
GRAVITY trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2017-05-16T18:28:13 (UT)
  | now: 2024-11-23T03:41:53 (UT) muc05 QC pipeline: gravity-1.0.2 (installed 2017-03-23)
same group: dark_low dark_med dark_high   bpix+gain_low bpix+gain_med bpix+gain_high acqcam
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
GRAVITY news:  
Report news:
DATE*: [?] 2024-11-16
report | NLT
2024-11-17
report | NLT
2024-11-18
report | NLT
2024-11-19
report | NLT
2024-11-20
report | NLT
2024-11-21
report | NLT
2024-11-22
report | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw no CAL raw files
P...
CAL product quality:
products products products products products products no CAL products
 
transfer   ngas [?]

scores&comments | history ... | plot tutorial ... | contact               [?]
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN -110 ADU VAL-250,-80 94 darklevel_ft this | last_yr | all mean dark level, Fringe Tracker (FT) detector, SPLIT polarimetry, DIT=0.00085 sec
2 QC1DB MEDIAN 11.3 ADU VAL10,15 94 biasavg_sc this | last_yr | all mean bias level, Science Camera (SC) detector, SPLIT polarimetry, DIT=0.3 sec
3 QC1DB MEDIAN -104 ADU VAL-250,-80 69 darklevel_ft this | last_yr | all mean dark level, Fringe Tracker (FT) detector, COMBINED polarimetry, DIT=0.00085 sec
4 QC1DB MEDIAN 20.7 ADU VAL20,25 69 biasavg_sc this | last_yr | all mean bias level, Science Camera (SC) detector, COMBINED polarimetry, DIT=0.3 sec
5 QC1DB MEDIAN 12.8 ADU VAL12,35 94 darkrms_ft this | last_yr | all RMS of dark, Fringe Tracker (FT) detector, SPLIT polarimetry, DIT=0.00085 sec
6 QC1DB MEDIAN 5.19 ADU VAL4.5,6.0 94 darkrms_sc this | last_yr | all RMS of dark, Science Camera (SC) detector, SPLIT polarimetry, DIT=0.3 sec
7 QC1DB MEDIAN 12.7 ADU VAL12,35 69 darkrms_ft this | last_yr | all RMS of dark, Fringe Tracker (FT) detector, COMBINED polarimetry, DIT=0.00085 sec
8 QC1DB MEDIAN 5.38 ADU VAL4.5,6.0 69 darkrms_sc this | last_yr | all RMS of dark, Science Camera (SC) detector, COMBINED polarimetry, DIT=0.3 sec
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 3|
data source:gravity_dark
(QC1 database)
dataset:darklevel_ft
median:-110ADU
fixed thresholds:-250...-80ADU
N_data plotted:94
[click on plot for closeup]
Plot 2
scores: 3|
data source:gravity_dark
(QC1 database)
dataset:biasavg_sc
median:11.3ADU
fixed thresholds:10...15ADU
N_data plotted:94
[click on plot for closeup]
Plot 3
scores: 7|
data source:gravity_dark
(QC1 database)
dataset:darklevel_ft
median:-104ADU
fixed thresholds:-250...-80ADU
N_data plotted:69
[click on plot for closeup]
Plot 4
scores: 7|
data source:gravity_dark
(QC1 database)
dataset:biasavg_sc
median:20.7ADU
fixed thresholds:20...25ADU
N_data plotted:69
[click on plot for closeup]
Plot 5
scores: 3|
data source:gravity_dark
(QC1 database)
dataset:darkrms_ft
median:12.8ADU
fixed thresholds:12...35ADU
N_data plotted:94
[click on plot for closeup]
Plot 6
scores: 3|
data source:gravity_dark
(QC1 database)
dataset:darkrms_sc
median:5.19ADU
fixed thresholds:4.5...6.0ADU
N_data plotted:94
[click on plot for closeup]
Plot 7
scores: 7|
data source:gravity_dark
(QC1 database)
dataset:darkrms_ft
median:12.7ADU
fixed thresholds:12...35ADU
N_data plotted:69
[click on plot for closeup]
Plot 8
scores: 7|
data source:gravity_dark
(QC1 database)
dataset:darkrms_sc
median:5.38ADU
fixed thresholds:4.5...6.0ADU
N_data plotted:69
[click on plot for closeup]
This plot

Trending plot for dark exposures of the Fringe Tracker (FT) and Science Camera (SC) detectors. Plotted are the dark level (FT), the bias level (SC), and the RMS of the master darks (both). The dark level is determined from the master dark whereas the bias is determined from the input raw frames and subtracted when creating the master in case of the SC detector.

The SC detector features a windowed read-out that depends on the spectral resolution (low, medium, or high) and on the polarimetry mode (splitted or combined). This plot shows QC parameters from darks with the read-out for MEDIUM resolution data. DITs are 0.3 sec (SC) and 0.00085 sec (FT), respectively.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up