mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
FLAMES/GIRAFFE:
lamps: stability
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
GIRAFFE trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-11-22T16:35:21 (UT)
  | now: 2024-11-22T18:51:41 (UT) muc02 QC pipeline: giraf-2.16.14 (installed 2024-10-16)
same group: SIM-lamp FF-lamp ARC-lamp   NAS-lamp
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
GIRAFFE news:  
Report news: 2023-01-22 Lamp ("Tungsten") was replaced and adjusted around Jan 22 2023, hence the bad values around that day. It took a while to sort out, but the red flag should vanish in the next week, once all settings have been re-taken [sciops-dsc@eso.org]
DATE*: [?] 2024-11-15
report | NLT
2024-11-16
report | NLT
2024-11-17
report | NLT
2024-11-18
report | NLT
2024-11-19
report | NLT
2024-11-20
report | NLT
2024-11-21
report | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw raw
P...
CAL product quality:
products products products products products products products
 
transfer   ngas [?]

scores&comments | FULL | history ... | plot tutorial ... | contact | HC_docu... | quality events...                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
- efficiency curves for Medusa settings
- reference flats
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 154 counts/sec VAL80,290 40 flux this | last_yr | all flux of (robotic) FF lamp, L543.1 setting; Medusa1
2 +QC1DB MEDIAN 148 counts/sec VAL80,290 25 flux this | last_yr | all as plot 1, Medusa2
3 QC1DB MEDIAN 76.5 counts/sec VAL18,170 12 flux this | last_yr | all as plot 1, IFU1
4 QC1DB MEDIAN 79.3 counts/sec VAL18,170 13 flux this | last_yr | all as plot 1, IFU2
5 QC1DB MEDIAN 41.4 counts/sec VAL10,90 14 flux this | last_yr | all as plot1 , Argus
6 QC1DB none   counts/sec none  40 flux this | last_yr | all as plot 1, combining all L543.1 flats; Medusa1
6 +QC1DB none   counts/sec none  25 flux this | last_yr | all     Medusa2
6 QC1DB none   counts/sec none  12 flux this | last_yr | all     IFU1
6 QC1DB none   counts/sec none  13 flux this | last_yr | all     IFU2
6 QC1DB none   counts/sec none  14 flux this | last_yr | all     Argus
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 3|
data source:giraffe_flat
(QC1 database)
dataset:flux
median:154counts/sec
fixed thresholds:80...290counts/sec
N_data plotted:40
[click plot for closeup]
Plot 2
scores: 3|
data source:giraffe_flat
(QC1 database)
dataset:flux+
median:148counts/sec
fixed thresholds:80...290counts/sec
N_data plotted:25
[click plot for closeup]
Plot 3
scores: 2|
data source:giraffe_flat
(QC1 database)
dataset:flux
median:76.5counts/sec
fixed thresholds:18...170counts/sec
N_data plotted:12
[click plot for closeup]
Plot 4
scores: 2|
data source:giraffe_flat
(QC1 database)
dataset:flux
median:79.3counts/sec
fixed thresholds:18...170counts/sec
N_data plotted:13
[click plot for closeup]
Plot 5
scores: 2|
data source:giraffe_flat
(QC1 database)
dataset:flux
median:41.4counts/sec
fixed thresholds:10...90counts/sec
N_data plotted:14
[click plot for closeup]
Plot 6
scores: not implemented
data source:giraffe_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
flux
average:none
thresholds:none
N_data plotted:40
[click plot for closeup]
This plot

This is the Health Check plot for the stability of the GIRAFFE flat field calibration lamps. The plot is based on ROBOTIC flats (flattype=ROB) taken several times a week. All 5 fibre systems (on two plates) are illuminated by the same flat lamp. Since June 2006, the HC flats are measured for all five slit systems. Before that date, they were taken for Medusa1 and Medusa2 only.

The average flux is obtained by dividing the average signal in the first raw file by the EXPTIME. Any effect showing up in only one fibre system is likely to be caused by that fibre system. Effects related to the lamp should be visible in any fibre system in the same way.

Find more tutorial information here.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up