mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
FLAMES/GIRAFFE:
system efficiency
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
GIRAFFE trending system: FULL report
HELP USERS-GUIDE MORE
Last update: 2024-11-13T04:40:27 (UT)
  | now: 2024-11-13T11:30:15 (UT) muc02 QC pipeline: giraf-2.16.14 (installed 2024-10-16)
same group: Efficiency per_WLN
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
GIRAFFE news:  
Report news: 2021-10-19 signal for the STD was low [qc_flames@eso.org]
DATE*: [?] 2024-11-06
no data | NLT
2024-11-07
no data | NLT
2024-11-08
no data | NLT
2024-11-09
report | NLT
2024-11-10
report | NLT
2024-11-11
report | NLT
2024-11-12
report | NLT
r...
Raw CAL displays:
no CAL raw files no CAL raw files no CAL raw files no CAL raw files raw raw raw
P...
CAL product quality:
no CAL products no CAL products no CAL products no CAL products products products products
 
transfer   ngas [?]

scores&comments | history: all data plotted here | plot tutorial ... | contact | HC_docu... | quality events...               [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Marked green: current ETC values.
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEAN 0.0326 fraction VAL0.015,0.07 182 mean_effic all (full time range) Efficiency at central wavelength (Argus, HR grating): H447.1A
2 QC1DB MEAN 0.0423 fraction VAL0.03,0.10 188 mean_effic all (full time range) H651.5B
3 QC1DB MEAN 0.256 fraction VAL0.11,0.40 174 mean_effic all (full time range) H875.7
4 QC1DB MEAN 0.0399 fraction VAL0.02,0.07 208 mean_effic all (full time range) Argus, LR grating: L427.2
5 QC1DB MEAN 0.0519 fraction VAL0.03,0.09 204 mean_effic all (full time range) L543.1
6 QC1DB MEAN 0.135 fraction VAL0.04,0.20 225 mean_effic all (full time range) L881.7
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 1|
data source:giraffe_std
(QC1 database)
dataset:mean_effic
mean:0.0326fraction
fixed thresholds:0.015...0.07fraction
N_data plotted:182
[click plot for closeup]
Plot 2
scores: 1|
data source:giraffe_std
(QC1 database)
dataset:mean_effic
mean:0.0423fraction
fixed thresholds:0.03...0.1fraction
N_data plotted:188
[click plot for closeup]
Plot 3
scores: 1|
data source:giraffe_std
(QC1 database)
dataset:mean_effic
mean:0.256fraction
fixed thresholds:0.11...0.4fraction
N_data plotted:174
[click plot for closeup]
Plot 4
scores: 1|
data source:giraffe_std
(QC1 database)
dataset:mean_effic
mean:0.0399fraction
fixed thresholds:0.02...0.07fraction
N_data plotted:208
[click plot for closeup]
Plot 5
scores: no current data
data source:giraffe_std
(QC1 database)
dataset:mean_effic
mean:0.0519fraction
fixed thresholds:0.03...0.09fraction
N_data plotted:204
[click plot for closeup]
Plot 6
scores: 1|
data source:giraffe_std
(QC1 database)
dataset:mean_effic
mean:0.135fraction
fixed thresholds:0.04...0.2fraction
N_data plotted:225
[click plot for closeup]
This plot

This is the Health Check plot to monitor the system efficiency of GIRAFFE. It is based on measurements of a flux standard star with the Argus IFU under photometric conditions. There are three settings taken with the low-resolution grating (L427.2, L543.1, L881.7) and three more with the high-resolution grating (H447.1A, H651.5B, H875.7). These data are taken approximately every month. Find a chromatic plot of these values here.

They are calibrated here with the robotic lamp flats. The flats are used only for the localization solution (required for the extraction). Contrary to the usual reduction, the extracted flat-field is not divided into the standard star spectrum, in order not to corrupt the efficiency calculation.

We display the QC1 parameter mean_effic which is the ratio of detected photons and the incident photons. The signal from all fibres (excluding SKY and SIMCAL) is combined and averaged.

For reference we also plot the current ETC values at the reference wavelength of the setting. Check out the ETC for more.
 


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up