mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
CRIRES:
wave
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
CRIRES trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2014-07-22T11:33:36 (UT)
  | now: 2024-11-23T07:08:55 (UT) muc01 QC pipeline: crire-2.3.2 (installed 2014-03-18)
same group: xcorr ThAr N2O DISP RESOL   AVG
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
CRIRES news:
Report news:
DATE*: [?] 2024-11-16
report | NLT
2024-11-17
report | NLT
2024-11-18
report | NLT
2024-11-19
report | NLT
2024-11-20
report | NLT
2024-11-21
report | NLT
2024-11-22
report | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw no CAL raw files
P...
CAL product quality:
products products products products products products no CAL products
  no OPSLOG data  

scores&comments | FULL | history ... | plot tutorial ... | contact                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB no MEDIAN 1083.785 nm OFF± .01 87 lambda_cent this | last_yr | all center wavelength on chip 2, WREF=1090.4
2 QC1DB no MEDIAN 1090.467 nm OFF± .01 87 lambda_cent this | last_yr | all center wavelength on chip 3, WREF=1090.4
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: OPSLOG data are included
Plot 1
scores: 8|
data source:crires_wave
(QC1 database)
dataset:lambda_cent
median:1083.785nm
rel. thresholds: 1083.7754...1083.7954nm
N_data plotted:87
[click on plot for closeup]
Plot 2
scores: 8|
data source:crires_wave
(QC1 database)
dataset:lambda_cent
median:1090.467nm
rel. thresholds: 1090.456675...1090.476675nm
N_data plotted:87
[click on plot for closeup]
This plot

Wavelength calibration: trending of central wavelength.

The central wavelength is monitored for the two daily health check settings with reference wavelengths (on chip 3) of 1090.4nm using ThAr lamp and 4092.9nm using N2O absorption cell. This plot shows the results with ThAr.

The center wavelength should be stable on each chip. A shift of 0.01 nm corresponds to a shift of 2 pixels. The variation of the central wavelength should be less than about 5 pixels. Chips 1 and 4 are not fully exposed due to the setting of the intermediate slit. Pipeline results for these chips are not plotted.  


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up