mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
CRIRES:
flat
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
CRIRES trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2014-07-22T11:34:23 (UT)
  | now: 2024-11-13T10:17:53 (UT) muc01 QC pipeline: crire-2.3.2 (installed 2014-03-18)
same group: flat Hal_IR   ThAr
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
CRIRES news:
Report news:
DATE*: [?] 2024-11-06
report | NLT
2024-11-07
report | NLT
2024-11-08
report | NLT
2024-11-09
report | NLT
2024-11-10
report | NLT
2024-11-11
report | NLT
2024-11-12
no data | NLT
r...
Raw CAL displays:
no CAL raw files no CAL raw files no CAL raw files raw raw raw no CAL raw files
P...
CAL product quality:
no CAL products no CAL products no CAL products products products products no CAL products
  last OPSLOG data:
2014-07-16*

last files ...
[an error occurred while processing this directive]
monitor

scores&comments | history ... | plot tutorial ... | contact                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB yes (1) MEDIAN 2.51e+03 ADU/sec VAL800,5000 87 mean_flux this | last_yr | all Halogen lamp, chip 1, WREF=2199.1
2 QC1DB yes (1) MEDIAN 2.49e+03 ADU/sec VAL800,4500 87 mean_flux this | last_yr | all Halogen lamp, chip 2, WREF=2199.1
3 QC1DB yes (1) MEDIAN 1.72e+03 ADU/sec VAL500,3500 87 mean_flux this | last_yr | all Halogen lamp, chip 3, WREF=2199.1
4 QC1DB yes (1) MEDIAN 1.25e+03 ADU/sec VAL300,2500 87 mean_flux this | last_yr | all Halogen lamp, chip 4, WREF=2199.1
5 QC1DB yes (1) MEDIAN 136 ADU/sec none  90 mean_flux this | last_yr | all IR emitter, chip 1, WREF=2199.1
6 QC1DB yes (1) MEDIAN 137 ADU/sec none  91 mean_flux this | last_yr | all IR emitter, chip 2, WREF=2199.1
7 QC1DB yes (1) MEDIAN 97.7 ADU/sec none  91 mean_flux this | last_yr | all IR emitter, chip 3, WREF=2199.1
8 QC1DB yes (1) MEDIAN 73.9 ADU/sec none  90 mean_flux this | last_yr | all IR emitter, chip 4, WREF=2199.1
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: OPSLOG data are included
Plot 1
scores: 9|
data source:crires_flat
(QC1 database)
dataset:mean_flux
median:2.51e+03ADU/sec
fixed thresholds:800...5000ADU/sec
N_data plotted:87
[click on plot for closeup]
Plot 2
scores: 9|
data source:crires_flat
(QC1 database)
dataset:mean_flux
median:2.49e+03ADU/sec
fixed thresholds:800...4500ADU/sec
N_data plotted:87
[click on plot for closeup]
Plot 3
scores: 9|
data source:crires_flat
(QC1 database)
dataset:mean_flux
median:1.72e+03ADU/sec
fixed thresholds:500...3500ADU/sec
N_data plotted:87
[click on plot for closeup]
Plot 4
scores: 9|
data source:crires_flat
(QC1 database)
dataset:mean_flux
median:1.25e+03ADU/sec
fixed thresholds:300...2500ADU/sec
N_data plotted:87
[click on plot for closeup]
Plot 5
scores: not implemented
data source:crires_flat
(QC1 database)
dataset:mean_flux
median:136ADU/sec
thresholds:none 
N_data plotted:90
[click on plot for closeup]
Plot 6
scores: not implemented
data source:crires_flat
(QC1 database)
dataset:mean_flux
median:137ADU/sec
thresholds:none 
N_data plotted:91
[click on plot for closeup]
Plot 7
scores: not implemented
data source:crires_flat
(QC1 database)
dataset:mean_flux
median:97.7ADU/sec
thresholds:none 
N_data plotted:91
[click on plot for closeup]
Plot 8
scores: not implemented
data source:crires_flat
(QC1 database)
dataset:mean_flux
median:73.9ADU/sec
thresholds:none 
N_data plotted:90
[click on plot for closeup]
This plot

Trending plots monitoring the flux in flat-fields taken with the Halogen lamp or with the IR emitter.

These flats are measured daily with a reference wavelength of 2199.1 nm. The exposures are taken for QC purposes; only the Halogen lamp is also used in other flat-field measurements that are intended for science calibration.  


General information

Click on any of the plots to see a close-up version.

If applicable, the latest values from Paranal ops logs are plotted as blue dots. More information about those values, including a download option, is available in the OPS column.

The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up