mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
XSHOOTER:
lamps: arc & flat
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
XSHOOTER trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-10-30T22:41:55 (UT)
  | now: 2024-10-31T06:06:55 (UT) muc02 QC pipeline: xshoo-3.6.5a (installed 2023-09-18)
same group: flat_HC_IFU slice_rms   flat_IFU flat_SLIT arc_lamp D2_spectrum QTH_spectrum SLIT stdev
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
XSHOOTER news:  
Report news:
DATE*: [?] 2024-10-24
report | NLT
2024-10-25
report | NLT
2024-10-26
report | NLT
2024-10-27
no data | NLT
2024-10-28
no data | NLT
2024-10-29
no data | NLT
2024-10-30
no data | NLT
r...
Raw CAL displays:
raw raw raw no CAL raw files no CAL raw files no CAL raw files no CAL raw files
P...
CAL product quality:
products products products no CAL products no CAL products no CAL products no CAL products
 
transfer   ngas [?]

scores&comments | history ... | plot tutorial ... | contact | HC_docu... | quality events...                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 0.279 ADU OFF± 0.03 199 slice2_rms this | last_yr | all rms over three slice (of a red or a blue echelle order) window counts
2 QC1DB MEDIAN 0.171 ADU OFF± 0.02 199 slice_rms this | last_yr | all rms over three slice (of a red or a blue echelle order) window counts
3 QC1DB none   NONE none  546 ambient_humidity this | last_yr | all relative humidity (extracted form flexure compensation frames
4 QC1DB MEDIAN -0.0379 ADU OFF± 0.04 199 slice2_s_rms this | last_yr | all relative difference of the two outer slices (of a red or a blue echelle order) window counts
5 QC1DB MEDIAN -0.023 ADU OFF± 0.03 199 slice_s_rms this | last_yr | all relative difference of the two outer slices (of a red or a blue echelle order) window counts
6 oQC1DB none   ratio none  15 flux_order_ratio this | last_yr | all counts ratio between the blue and the red ecehlle order (IFU flats only; 5 raw frames)
6 QC1DB none   ratio none  199 flux_order_ratio this | last_yr | all counts ratio between the blue and the red echelle order (IFU HC flats only)
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 14|
data source:xshooter_flat
(QC1 database)
dataset:slice2_rms
median:0.279ADU
rel. thresholds: 0.2490335...0.3090335ADU
N_data plotted:199
[click plot for closeup]
Plot 2
scores: 14|
data source:xshooter_flat
(QC1 database)
dataset:slice_rms
median:0.171ADU
rel. thresholds: 0.151376...0.191376ADU
N_data plotted:199
[click plot for closeup]
Plot 3
scores: not implemented
data source:xshooter_flexure
(QC1 database)
dataset:ambient_humidity
average:none
thresholds:none
N_data plotted:546
[click plot for closeup]
Plot 4
scores: 14|
data source:xshooter_flat
(QC1 database)
dataset:slice2_s_rms
median:-0.0379ADU
rel. thresholds: -0.0779355...0.0020645ADU
N_data plotted:199
[click plot for closeup]
Plot 5
scores: 14|
data source:xshooter_flat
(QC1 database)
dataset:slice_s_rms
median:-0.023ADU
rel. thresholds: -0.052981...0.007019ADU
N_data plotted:199
[click plot for closeup]
Plot 6
scores: not implemented
data source:xshooter_flat
(QC1 database)
dataset:
(numbers below apply to this dataset)
flux_order_ratioo
average:none
thresholds:none
N_data plotted:15
[click plot for closeup]
This plot

This plot shows two QC parameters measuring the illumination distribution within IFU FLATs. Both kind of IFU flats contribute:

  • IFU health check flats generated from one (VIS) or two (NIR/UVB) raw flats and
  • IFU master flats generated from five (VIS) or ten (UVB/NIR) raw flats, acquired to calibrate science data.
 
General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up