mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
OMEGACAM:
bias
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
OMEGACAM trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-11-04T16:00:14 (UT)
  | now: 2024-11-05T00:20:17 (UT) muc06 QC pipeline: omega-1.1.9 (installed 2024-07-29)
same group: BIAS_level BIAS_readnoise RON_select BIAS_select BIAS_hotpixels
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
OMEGACAM news:  
Report news:
DATE*: [?] 2024-10-29
report | NLT
2024-10-30
report | NLT
2024-10-31
report | NLT
2024-11-01
report | NLT
2024-11-02
report | NLT
2024-11-03
report | NLT
2024-11-04
report | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw raw
P...
CAL product quality:
products products products products products products products
 
transfer   ngas [?]

scores&comments | history ... | plot tutorial ... | contact | HC_docu... | quality events...               [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
    OmegaCAM temperature probe keywords no longer exist. Until fixed, temp_AVG can be ignored.
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 2.16 ADU VAL1.5,2.7 191 qc_ron this | last_yr | all BIAS readnoise level averaged over all 32 detectors (ADU)
2 QC1DB MEDIAN 2.11 ADU none  6112 qc_ron this | last_yr | all BIAS readnoise level for each individual detector (ADU)
3 QC1DB MEDIAN 0.221 ADU VAL0.,0.5 191 qc_ron this | last_yr | all BIAS readnoise level rms averaged over all 32 detectors (ADU)
4 LOCAL none   none none  6074 det_id n/a score per detector:
GREEN = within the thresholds
4 LOCAL none   none none  38 det_id n/a score per detector:
RED = beyond the thresholds
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 19|
data source:omegacam_readnoise
(QC1 database)
dataset:qc_ron
median:2.16ADU
fixed thresholds:1.5...2.7ADU
N_data plotted:191
[click plot for closeup]
Plot 2
scores: not implemented
data source:omegacam_readnoise
(QC1 database)
dataset:qc_ron
median:2.11ADU
thresholds:none 
N_data plotted:6112
[click plot for closeup]
Plot 3
scores: 19|
data source:omegacam_readnoise
(QC1 database)
dataset:qc_ron
median:0.221ADU
fixed thresholds:0.0...0.5ADU
N_data plotted:191
[click plot for closeup]
Plot 4
scores: not implemented
data source:local data source
dataset:
(numbers below apply to this dataset)
det_id
average:none
thresholds:none
N_data plotted:6074
[click plot for closeup]
This plot

Trending plots for the bias read-out noise.

QC parameters are derived from each of 32 detectors of OmegaCAM BIAS frames.

DB Name: omegacam_readnoise
Fits header: QC.READNOISE
QC1 DB: qc_ron / qc_temp_mosaic_plate

Description:
The bias read-out noise as measured from the pipeline product frame READNOISE_ADU

Setup:
BIN.X / BIN.Y / READ.MODE / READ.SPEED = 1/1/normal/normal

These Health Check plots describe:
Fig. 1 readnoise_AVG: The bias read-out noise averaged over all 32 detectors (in ADU). This plot is sensitive to any changes common to all detectors.
Fig. 2 readnoise_ALL: The bias read-out noise for each individual detector (in ADU). This plot is sensitive changes in individual detectors.
Fig. 3 readnoise_RMS: The rms of the bias read-out noise averaged over all 32 detectors (in ADU).
Fig. 4 score_outliers: Scores of the bias readnoise as measured for each of the 32 detectors.
Green = within the defined thresholds.
Red = beyond the defined thresholds.


OmegaCAM detector plane layout (detector_ID locations):
OmegaCAM detector layout

 
General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up