NACO: Dark
|
HC PLOTS |
RON, dark level: |
|
hot pixel, cold pixel: |
|
QC1 database (advanced users):
browse |
plot
|
A number of DARK frames is taken within the NACO daytime calibrations in case
NACO science observations have been acquired in the night. Each DARK consists
of 3 raw frames obtained with the instrumental setting matching the one used
for science observations. The parameters matched are:
- DET.DIT - detector integration time
- DET.NDIT - number of sub-integration
- INS.OPTI7.NAME - camera (e.g. S17, L54)
- DET.NCORRS.NAME - detector read-out mode
- DET.MODE.NAME - detector mode
In addition, each day, 3 dark frames are taken for the health check purpose,
always with the same setup:
- DET.DIT=1.7927 s, (since 2015: 1.7892s) DET.NDIT=1, INS.OPTI7.NAME=S27, DET.NCORRS.NAME=FowlerNsamp, DET.MODE.NAME=HighSensitivity
- DET.DIT=0.5 s, DET.NDIT=100, INS.OPTI7.NAME=S27, DET.NCORRS.NAME=Double_RdRstRd, DET.MODE.NAME=HighSensitivity
- DET.DIT=300 s, DET.NDIT=1, INS.OPTI7.NAME=S27, DET.NCORRS.NAME=FowlerNsamp, DET.MODE.NAME=HighSensitivity
They are utilized to monitor performance of the NACO detector.
The following images are from 2016-11-28.
|
DARK frame - DIT=0.5s, Double_RdRstRd, HighSensitivity
This is an example of the master calibration DARK
frame. It is a pipeline product of the "naco_img_dark" recipe.
The horizontal stripes on the edges of the detector are features characteristic to the Double_RdRstRd read-out mode.
Note the dead columns (every 8th is affected) in the lower left quadrant. The bright dot at the upper edge and the upper right corner are known bad pixel areas.
|
|
DARK frame - DIT=1.7s, FowlerNsamp, HighSensitivity This is an example of the master calibration DARK
frame. It is a pipeline product of the "naco_img_dark" recipe.
The bright, diagonal stripes are associated with the "light leakage". They become
prominent structures when darks are taken with long DITs.
|
|
DARK frame - DIT=300s, FowlerNsamp, HighSensitivity This is an example of the master
calibration DARK frame. It is a pipeline product of the "naco_img_dark" recipe.
The bright and wide diagonal stripes are associated with the "light leakage". They become prominent structures
when darks are taken with long DITs. The bright stripes are variable:
the pattern occurs at a slightly different position each time the camera is moved.
|
|
DARK frame - DIT=0.18s, Uncorr, HighWellDepth This is an
example of the master calibration DARK frame.
It is a pipeline product of the "naco_img_dark" recipe. The dark level is (please note that uncorrelated read mode means a reset-read sequence) at -1800 ADU
|
In addition to the QC parameters calculated by the NACO dark pipeline recipe,
several QC parameters are calculated by QC procedures. They all are stored in
the NACO QC1 parameter database.
The NACO detector can be operated in normal (DET.FRAM.TYPE=INT) and in cube
mode (DET.FRAM.TYPE=CUBE2). In cube mode, the detector adds two more rows at
the upper rim, that the read-out size is 1024x1026 pixel instead of the
1024x1024 pixel in normal read mode. Furthermore the detector can be operated
in any used-defined window read out mode. The QC monitoring covers:
- normal mode (DET.FRAM.TYPE=INT), full readout (NAXIS1 x NAXIS2 = 10124 x 1024)
- darks for M-band pre-defined 512 x 512 pixel window readout and 512 x 514 pixel window (cube mode)
Window read out options other than 512 x 512 and 512 x 514 are not pipeline
supported and are not subject of QC checks.
In addition to the trending plots which show the evolution of the QC
parameter with time, we provide a diagnostic plot, which shows the relation
between the dark level and the exposure time
to document the reduction of dark current since 2015.
The operationally most critical detector characteristics derived from dark
calibration frames are monitored by the following four QC parameters:
Readout noise
The read out noise is the signal scatter when no light enters the
detector. It is a basic property of the detector (of each pixel) and
can be derived by multiple read outs. We monitor the NACO detector read out
in all detector read modes
and DIT.
QC1_parameters
FITS key |
QC1 database: table, name |
definition |
class* |
HC_plot** |
more docu |
QC.RON1 | naco_dark..dark_ron12 | RON (from dark-ron recipe raw2-raw1) | HC | | [docuSys coming] |
*Class: KPI - instrument performance; HC - instrument health; CAL - calibration quality; ENG - engineering parameter
**There might be more than one. |
Trending
The readout noise is plotted in boxes 1,2 and 3 of
the trending
plot.
Scoring&thresholds Readout noise
Scoring thresholds have been chosen to be slightly
more tolerant than 3 sigma, where sigma is the scatter
of the data RON values acquired after 2016-01.
History
The stability of the readout noise since start of operations in 2004 is documented
here
where the green points refer to values of the replaced detector from 2015 on.
The detector readout noise for DIT=300s, Fowler sampling read mode is highly contaminated by the light leakage.
Algorithm Readout noise
The readout noise is calculated from the 2nd-1st raw dark frame;
it is a median of the standard deviations measured from 100 4x4 windows on the difference frame, multiplied by sqrt(NDIT/2)
Dark level
The dark level is the mean signal achieved from a detector when no light
enters the detector. In double correlated read mode, when two signals are subtracted from each other,
the dark level should be around zero for small DIT and should linearly increase with time for longer exposure times.
In uncorrelated read mode, the dark level is the configured bias level (for short DIT) and can be negative ( ~ -3000 ADU).
QC1_parameters
FITS key |
QC1 database: table, name |
definition |
class* |
HC_plot** |
more docu |
QC.DARKMED | naco_dark..dark_median | median (from stcube recipe) | HC | | [docuSys coming] |
*Class: KPI - instrument performance; HC - instrument health; CAL - calibration quality; ENG - engineering parameter
**There might be more than one. |
Trending
The dark level is plotted in boxes 4,5 and 6 of
the trending
plot.
Scoring&thresholds Dark level
The dark level is scored. Thresholds have been derived from the scatter in the dark level from data acquired after 2015-06.
History
- 2015-01-01: detector Raytheon_Aladdin3_InSb replaced by AladdinIRD02
- The stability of the dark level since start of operations in 2004 is documented
here
where the green points refer to values of the replaced detector from 2015 on.
The detector dark level for DIT=300s, Fowler sampling read mode is highly contaminated by the light leakage.
Algorithm Dark level
The median dark level of the master dark excludes the region of a 100 pixel wide stripe at the rim of the detector.
Hot and cold pixel
The number of hot and cold pixel are derived from master dark products.
- hot pixel are defined as master dark pixel with more than TH ADU above the dark level.
- cold pixel are defined as master dark pixel with less than TC ADU below the dark level.
Since 2015, when the NACO is operated with a new (the former ISAAC LW-arm detector), TH=14 and TC=10 are used.
QC1_parameters
FITS key |
QC1 database: table, name |
definition |
class* |
HC_plot** |
more docu |
QC.NBHOTPIX | naco_dark..num_hot | number of hot pixels | HC | | [docuSys coming] |
QC.NBCOLDPIX | naco_dark..num_cold | number of cold pixels | HC | | [docuSys coming] |
*Class: KPI - instrument performance; HC - instrument health; CAL - calibration quality; ENG - engineering parameter
**There might be more than one. |
Trending
The definition of the hot and cold pixel makes only sense,
when the master dark is flat, which is fulfilled for the
short exposure time Uncorr and Double read modes. Since
2015, the known light leakage produces even brighter
diagonal stripes than before 2015. Therefore the hot
pixel registered in DIT=300sec Fowler sampling are pixel
contaminated by the light leakage.
Scoring&thresholds Hot and cold pixel
These detector engineering QC parameters are not scored. Hot
and cold pixel are not propagated in the data reduction cascade.
History
before 2015, another detector was in operation, hot and
cold pixel have been derived using the pipeline default
parameters TH=10 and TC=6.
Algorithm Hot and cold pixel
Count the number of pixel in the master dark with more then TH
ADU above the median level and less than TC ADU below the median dark level.
|