mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
MIDI:
wavelength calib
Other HC:
QC links:
[an error occurred while processing this directive]
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+R to enforce refresh of scores and dates]
MIDI trending system: HEALTH CHECK report
Last update: 2012-01-30T10:04:29 (UT)
| now: 2024-06-29T21:24:19 (UT)
HELP USERS-GUIDE MORE
same group: HS SP_INTERF SP_PHOT
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
The links to KMOS, UVES, and ESPRESSO have been moved to a new section (qcFlow operational) at the lower part of the navigation bar.
MIDI news:  
Report news:  
DATE*: [?] 2015-02-27
report | NLT
2015-02-28
report | NLT
2015-03-01
report | NLT
2015-03-02
report | NLT
2015-03-03
report | NLT
2015-03-04
report | NLT
2015-03-05
no data | NLT
P...
Product quality:
products products products products products products no products
  last OPSLOG data:
2012-01-29*

last files ...
[an error occurred while processing this directive]
monitor
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
- general comment The thresholds are set to the accuracies mentioned in the user manual:
0.2 microns for the PRISM
0.05 for the GRISM
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB yes (3) MEDIAN 8.93865 microns VAL8.839,9.039 58 ArIII_r1 this | last_yr | all ArIII_R1 ArIII_R1
2 QC1DB yes (3) MEDIAN 10.489 microns VAL10.389,10.589 58 SIV_r1 this | last_yr | all SIV_R1 SIV_R1
3 QC1DB yes (3) MEDIAN 12.72029 microns VAL12.62,12.82 58 NeII_r1 this | last_yr | all NeII_R1 NeII_R1
4 QC1DB yes (3) MEDIAN 8.93865 microns VAL8.839,9.039 58 ArIII_r4 this | last_yr | all ArIII_R4 ArIII_R4
5 QC1DB yes (3) MEDIAN 10.489 microns VAL10.389,10.589 58 SIV_r4 this | last_yr | all SIV_R4 SIV_R4
6 QC1DB yes (3) MEDIAN 12.72029 microns VAL12.62,12.82 58 NeII_r4 this | last_yr | all NeII_R4 NeII_R4
7 QC1DB yes (3) MEDIAN 8.95597 microns VAL8.931,8.981 58 ArIII_r1 this | last_yr | all ArIII_R1 ArIII_R1
8 QC1DB yes (3) MEDIAN 10.44284 microns VAL10.418,10.468 58 SIV_r1 this | last_yr | all SIV_R1 SIV_R1
9 QC1DB yes (3) MEDIAN 12.84069 microns VAL12.816,12.866 58 NeII_r1 this | last_yr | all NeII_R1 NeII_R1
10 QC1DB yes (3) MEDIAN 9.0256 microns VAL9.0006,9.0506 58 ArIII_r4 this | last_yr | all ArIII_R4 ArIII_R4
11 QC1DB yes (3) MEDIAN 10.489 microns VAL10.464,10.514 58 SIV_r4 this | last_yr | all SIV_R4 SIV_R4
12 QC1DB yes (3) MEDIAN 12.88677 microns VAL12.862,12.912 58 NeII_r4 this | last_yr | all NeII_R4 NeII_R4
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: to indicate that OPSLOG data are included
Plot 1
scores: 8|
data source:midi_wave
(QC1 database)
dataset:ArIII_r1
median:8.93865microns
fixed thresholds:8.839...9.039microns
N_data plotted:58
[click on plot for closeup]
Plot 2
scores: 8|
data source:midi_wave
(QC1 database)
dataset:SIV_r1
median:10.489microns
fixed thresholds:10.389...10.589microns
N_data plotted:58
[click on plot for closeup]
Plot 3
scores: 8|
data source:midi_wave
(QC1 database)
dataset:NeII_r1
median:12.72029microns
fixed thresholds:12.62...12.82microns
N_data plotted:58
[click on plot for closeup]
Plot 4
scores: 8|
data source:midi_wave
(QC1 database)
dataset:ArIII_r4
median:8.93865microns
fixed thresholds:8.839...9.039microns
N_data plotted:58
[click on plot for closeup]
Plot 5
scores: 8|
data source:midi_wave
(QC1 database)
dataset:SIV_r4
median:10.489microns
fixed thresholds:10.389...10.589microns
N_data plotted:58
[click on plot for closeup]
Plot 6
scores: 8|
data source:midi_wave
(QC1 database)
dataset:NeII_r4
median:12.72029microns
fixed thresholds:12.62...12.82microns
N_data plotted:58
[click on plot for closeup]
Plot 7
scores: 8|
data source:midi_wave
(QC1 database)
dataset:ArIII_r1
median:8.95597microns
fixed thresholds:8.931...8.981microns
N_data plotted:58
[click on plot for closeup]
Plot 8
scores: 8|
data source:midi_wave
(QC1 database)
dataset:SIV_r1
median:10.44284microns
fixed thresholds:10.418...10.468microns
N_data plotted:58
[click on plot for closeup]
Plot 9
scores: 8|
data source:midi_wave
(QC1 database)
dataset:NeII_r1
median:12.84069microns
fixed thresholds:12.816...12.866microns
N_data plotted:58
[click on plot for closeup]
Plot 10
scores: 8|
data source:midi_wave
(QC1 database)
dataset:ArIII_r4
median:9.0256microns
fixed thresholds:9.0006...9.0506microns
N_data plotted:58
[click on plot for closeup]
Plot 11
scores: 8|
data source:midi_wave
(QC1 database)
dataset:SIV_r4
median:10.489microns
fixed thresholds:10.464...10.514microns
N_data plotted:58
[click on plot for closeup]
Plot 12
scores: 8|
data source:midi_wave
(QC1 database)
dataset:NeII_r4
median:12.88677microns
fixed thresholds:12.862...12.912microns
N_data plotted:58
[click on plot for closeup]
This plot

The wavelength day calibration consists in using the MIDI heated blackscreen and taking 3 exposures through different narrow-band filters [NeII], [SIV] and [ArIII] (having central wavelengths accurately measured in the laboratory by the manufacturer). An extra exposure is taken through a polycarbonic foil that has several absorption lines in its N-band spectrum. Exposures without filters and with closed shutter are also taken to process the exposures with filters. From the spectra obtained through the filters a fit of the lambda(X) function is performed. The procedure is performed for each setup (PRISM or GRISM, and SCI_PHOT or HIGH_SENS). The MIDI HIGH_SENS mode uses 2 windows on the detector (R1 and R2). The MIDI SCI_PHOT mode uses 4 windows (R1 to R4) We monitor here the stability of the dispersion by plotting the wavelength of a spectral channel given by the lambda(X) function computed by the reduction of the calibration data. The selected spectral channels corresponds to the expected central wavelength of the three narrow-band filters.  


General information

Click on any of the plots to see a close-up version.

If applicable, the latest values from Paranal ops logs are plotted as blue dots. More information about those values, including a download option, is available in the OPS column.

The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up