mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
MIDI:
detector characteristics
Other HC:
QC links:
[an error occurred while processing this directive]
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+R to enforce refresh of scores and dates]
MIDI trending system: HEALTH CHECK report
Last update: 2012-01-30T08:51:53 (UT)
| now: 2024-06-26T18:15:33 (UT)
HELP USERS-GUIDE MORE
same group: RON LIN PRESS_TEMP
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
The links to KMOS, UVES, and ESPRESSO have been moved to a new section (qcFlow operational) at the lower part of the navigation bar.
MIDI news:  
Report news:  
DATE*: [?] 2015-02-27
report | NLT
2015-02-28
report | NLT
2015-03-01
report | NLT
2015-03-02
report | NLT
2015-03-03
report | NLT
2015-03-04
report | NLT
2015-03-05
no data | NLT
P...
Product quality:
products products products products products products no products
  last OPSLOG data:
2012-01-29*

last files ...
[an error occurred while processing this directive]
monitor
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
- general comment: Increase of the RON is often due to power cuts and/or high pressure (see pressure plot)
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB yes (3) none   ADU VAL10,15 60 median_masterron this | last_yr | all read-out noise of master frame (ADU)
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: to indicate that OPSLOG data are included
Plot 1
scores: 8|
data source:midi_detron
(QC1 database)
dataset:median_masterron
average:none
fixed thresholds:10.0...15.0ADU
N_data plotted:60
[click on plot for closeup]
This plot

The readout noise of MIDI is monitored by taking a full-frame exposure with a large number of frames, with the MIDI shutter closed and with a minimum DIT. The standard deviation of the pixel level over the frames is computed by the pipeline for each pixel, as well as the median of the standard deviation over all the pixels and all the frames, which is plotted here (expressed in detector ADUs).  


General information

Click on any of the plots to see a close-up version.

If applicable, the latest values from Paranal ops logs are plotted as blue dots. More information about those values, including a download option, is available in the OPS column.

The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up