mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
GRAVITY:
wavelength
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
GRAVITY trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2017-04-04T12:24:42 (UT)
  | now: 2024-08-17T02:15:40 (UT) muc05 QC pipeline: gravity-1.0.2 (installed 2017-03-23)
same group: refpos_low refpos_med refpos_high minmax_low minmax_med minmax_high   Ar_lamp_1 Ar_lamp_2
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
GRAVITY news:  
Report news:
DATE*: [?] 2024-08-10
report | NLT
2024-08-11
report | NLT
2024-08-12
report | NLT
2024-08-13
report | NLT
2024-08-14
report | NLT
2024-08-15
report | NLT
2024-08-16
report | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw no CAL raw files
P...
CAL product quality:
products products products products products products no CAL products
 
transfer   ngas [?]

scores&comments | history ... | plot tutorial ... | contact               [?]
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 2.34 microns none  73 maxwave_ft this | last_yr | all average coherence of P2VM, SPLIT polarisation, Fringe Tracker (FT) detector
2 QC1DB MEDIAN 2.47 microns none  73 maxwave_sc this | last_yr | all average coherence of P2VM, SPLIT polarisation, Science Camera (SC) detector
3 QC1DB MEDIAN 2.34 microns none  39 maxwave_ft this | last_yr | all average coherence of P2VM, COMBINED polarisation, Fringe Tracker (FT) detector
4 QC1DB MEDIAN 2.49 microns none  39 maxwave_sc this | last_yr | all average coherence of P2VM, COMBINED polarisation, Science Camera (SC) detector
5 QC1DB MEDIAN 2.01 microns none  73 minwave_ft this | last_yr | all phase RMS of P2VM, SPLIT polarisation, Fringe Tracker (FT) detector
6 QC1DB MEDIAN 1.97 microns none  73 minwave_sc this | last_yr | all phase RMS of P2VM, SPLIT polarisation, Science Camera (SC) detector
7 QC1DB MEDIAN 2 microns none  39 minwave_ft this | last_yr | all phase RMS of P2VM, COMBINED polarisation, Fringe Tracker (FT) detector
8 QC1DB MEDIAN 1.97 microns none  39 minwave_sc this | last_yr | all phase RMS of P2VM, COMBINED polarisation, Science Camera (SC) detector
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: not implemented
data source:gravity_p2vm
(QC1 database)
dataset:maxwave_ft
median:2.34microns
thresholds:none 
N_data plotted:73
[click on plot for closeup]
Plot 2
scores: not implemented
data source:gravity_p2vm
(QC1 database)
dataset:maxwave_sc
median:2.47microns
thresholds:none 
N_data plotted:73
[click on plot for closeup]
Plot 3
scores: not implemented
data source:gravity_p2vm
(QC1 database)
dataset:maxwave_ft
median:2.34microns
thresholds:none 
N_data plotted:39
[click on plot for closeup]
Plot 4
scores: not implemented
data source:gravity_p2vm
(QC1 database)
dataset:maxwave_sc
median:2.49microns
thresholds:none 
N_data plotted:39
[click on plot for closeup]
Plot 5
scores: not implemented
data source:gravity_p2vm
(QC1 database)
dataset:minwave_ft
median:2.01microns
thresholds:none 
N_data plotted:73
[click on plot for closeup]
Plot 6
scores: not implemented
data source:gravity_p2vm
(QC1 database)
dataset:minwave_sc
median:1.97microns
thresholds:none 
N_data plotted:73
[click on plot for closeup]
Plot 7
scores: not implemented
data source:gravity_p2vm
(QC1 database)
dataset:minwave_ft
median:2microns
thresholds:none 
N_data plotted:39
[click on plot for closeup]
Plot 8
scores: not implemented
data source:gravity_p2vm
(QC1 database)
dataset:minwave_sc
median:1.97microns
thresholds:none 
N_data plotted:39
[click on plot for closeup]
This plot

Trending plot for wavelength calibration.

The positions of two reference wavelengths (2.099 and 2.314 microns) on the detectors are trended for the lowest spectrum on the detector. Scoring thresholds have been set up provisionally to +/-2 pixels around the median (for SC high resolution) and +/-1 pixel for SC low and medium resolution.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up