NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
GRAVITY news:
Report news:
2024-12-13 The problem with the increasing bad pixels on the FT is still under investigation. To avoid the constant red flags, we increased the thresholds for the moment [lschmidt@eso.org]
Trending plot for the number of bad pixels of the Fringe Tracker (FT) and Science Camera
(SC) detectors. Values are determined from flat-field measurements.
General information
Click on any of the plots to see a close-up version.
The latest date is indicated on top of the plot.
If configured,
data points belonging to the latest date are specially marked
statistical averages are indicated by a solid line, and thresholds by broken lines
outliers are marked by a red asterisk. They are defined as data points outside the
threshold lines
"aliens" (= data points outside the plot Y limits) are marked by a red arrow (↑ or ↓)
you can download the data for each parameter set if the 'Data downloads' link shows up