mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
FLAMES/GIRAFFE:
detector: monitoring
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
GIRAFFE trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-12-21T16:33:27 (UT)
  | now: 2024-12-22T04:18:53 (UT) muc02 QC pipeline: giraf-2.16.14 (installed 2024-10-16)
same group: dark linearity&gain fp-noise contamination   correlations
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
GIRAFFE news:  
Report news:
DATE*: [?] 2024-12-15
report | NLT
2024-12-16
report | NLT
2024-12-17
report | NLT
2024-12-18
report | NLT
2024-12-19
report | NLT
2024-12-20
report | NLT
2024-12-21
report | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw no CAL raw files
P...
CAL product quality:
products products products products products products no CAL products
 
transfer   ngas [?]

scores&comments | history ... | plot tutorial ... | contact | HC_docu... | quality events...               [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 0.0163 fraction VAL0.010,0.020 12 fraction_sig_fp this | last_yr | all fixed-pattern noise, extracted from QC.FPN parameter (normal read mode).
Note: downloaded data: any exposure time!
2 QC1DB none   ADU none  276 sigma_fp this | last_yr | all measured fixed-pattern noise, with linear fit
2 QC1DB none   ADU none  235 sigma_ph this | last_yr | all measured photon noise, with square-root fit, values with mean_signal > 55000 suppressed
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 1|
data source:giraffe_ccd
(QC1 database)
dataset:fraction_sig_fp
median:0.0163fraction
fixed thresholds:0.01...0.02fraction
N_data plotted:12
[click plot for closeup]
Plot 2
scores: not implemented
data source:giraffe_ccd
(QC1 database)
dataset:
(numbers below apply to this dataset)
sigma_fp
average:none
thresholds:none
N_data plotted:276
[click plot for closeup]
This plot

This is the trending plot for the monitoring of the GIRAFFE detector noise parameters, derived from image flats.

Data for the old CCD (used until 2008-03) are complete. Due to issues with the pipeline recipe for detector monitoring, data for the new CCD have been calculated only since 2009-10.

These frames are taken approximately once per month, as a sequence with exposure time increasing from 0.5 sec to 220 sec. For each value, a set of 2 image flats is taken, just exposing the CCD without the fibre system.

For each pair, the difference frame is calculated from which the photon noise is determined. Then a raw frame is taken, shifted by one pixel in X and Y, and subtracted, in order to randomize the fixed-pattern noise. Finally the measured total noise and the photon noise are subtracted and give the fixed-pattern noise. It is trended as fraction of the mean signal in plot no. 1.

The second plot displays the whole sequence of image flats, with the measured photon noise on a square-root curve, and the fixed-pattern noise on a linear curve. The intersection of both fits marks the transition from the photon-noise to the fixed-pattern noise regime.

Find more information about GIRAFFE image flats frames here.

Other QC parameters related to detector properties (like dark, linearity, gain) are trended in the same group.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up