GIRAFFE trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-12-21T16:37:28 (UT)
  | now: 2024-12-22T03:16:56 (UT) muc02 QC pipeline: giraf-2.16.14 (installed 2024-10-16)
same group: rms N_lines resolution
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
GIRAFFE news:  
Report news: 2024-02-02 long-term trend in losing resolution (in all modes if you look at the full history) is known. This requires re-focusing the instrument, which will be done in due course. BHa [bhaeussl@eso.org]
DATE*: [?] 2024-12-15
report | NLT
2024-12-16
report | NLT
2024-12-17
report | NLT
2024-12-18
report | NLT
2024-12-19
report | NLT
2024-12-20
report | NLT
2024-12-21
no data | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw no CAL raw files
P...
CAL product quality:
products products products products products products no CAL products
 
transfer   ngas [?]

scores&comments | FULL | history ... | plot tutorial ... | contact | HC_docu... | quality events...                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
- reference arclamp files
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 3.01e+04 N VAL28000,34000 40 mean_R this | last_yr | all resolving power R (λ0/Δλ) H525.8B, Medusa1
2 xQC1DB MEDIAN 3.02e+04 N VAL28000,34000 27 mean_R this | last_yr | all as plot 1, Medusa2
3 QC1DB MEDIAN 4.63e+04 N VAL45000,55000 13 mean_R this | last_yr | all as plot 1, IFU1
4 QC1DB MEDIAN 4.59e+04 N VAL42000,55000 13 mean_R this | last_yr | all as plot 1, IFU2
5 QC1DB MEDIAN 4.84e+04 N VAL45000,55000 14 mean_R this | last_yr | all as plot 1, Argus
6 QC1DB MEDIAN 5.79e+03 N VAL5600,6200 37 mean_R this | last_yr | all resolving power R
L543.1, Medusa1
7 xQC1DB MEDIAN 5.79e+03 N VAL5600,6200 28 mean_R this | last_yr | all as plot 6, Medusa2
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 2|
data source:giraffe_wave
(QC1 database)
dataset:mean_R
median:3.01e+04N
fixed thresholds:28000...34000N
N_data plotted:40
[click plot for closeup]
Plot 2
scores: 2|
data source:giraffe_wave
(QC1 database)
dataset:mean_Rx
median:3.02e+04N
fixed thresholds:28000...34000N
N_data plotted:27
[click plot for closeup]
Plot 3
scores: 1|
data source:giraffe_wave
(QC1 database)
dataset:mean_R
median:4.63e+04N
fixed thresholds:45000...55000N
N_data plotted:13
[click plot for closeup]
Plot 4
scores: 2|
data source:giraffe_wave
(QC1 database)
dataset:mean_R
median:4.59e+04N
fixed thresholds:42000...55000N
N_data plotted:13
[click plot for closeup]
Plot 5
scores: 2|
data source:giraffe_wave
(QC1 database)
dataset:mean_R
median:4.84e+04N
fixed thresholds:45000...55000N
N_data plotted:14
[click plot for closeup]
Plot 6
scores: 2|
data source:giraffe_wave
(QC1 database)
dataset:mean_R
median:5.79e+03N
fixed thresholds:5600...6200N
N_data plotted:37
[click plot for closeup]
Plot 7
scores: 2|
data source:giraffe_wave
(QC1 database)
dataset:mean_Rx
median:5.79e+03N
fixed thresholds:5600...6200N
N_data plotted:28
[click plot for closeup]
This plot

This is the Health Check plot for monitoring the resolution of the GIRAFFE arclamp calibration data. We display here the resolving power R = λ0/Δλ. Δλ is the QC1 parameter mean_resol which is the mean of all accepted emission line widths. This plot monitors the HC data taken in high resolution (H525.8B) and low resolution (L543.1, Medusa1 and 2 only).

Up to 2008-01, these numbers are approximate. With pipeline version giraf-2.5 accurate numbers are provided. Only these should be taken for reference.

Check the RMS plot of this group for the typical rms of the dispersion solution, and the NLIN plot for the number of accepted lines.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up