IFU and ARGUS calibration frames
Note that for the fiber flats and arcs for the IFU modes, the appearence and counts are different depending on which plate (1 or 2) is used. ARGUS data only use plate 2.
Due to the large number of possible wavelength settings (22 HR settings
+ 8 LR settings for 3 slits, which makes [22+8]x3 = 90 settings in IFU and ARGUS),
only a selected few are displayed here. The counts should not exceed around
60,000 ADU on the GIRAFFE CCD. In the bluest settings, the counts may
be as low as 4,000 ADU, even using long integration times. Counts much
below this should be flagged, as they could indicate problems with the
lamp.
In the IFU modes, each of the bundles is made up of many fibers,
as is apparent by looking at the cross-cuts on the example images. The
cuts are very different when plate 1 and plate 2 are compared.
Note that for ARGUS, only plate 2 is used.