VIMOS MOS DFOS_OPS calibration cascade

DFOS_OPS | CALCHECKER | CALSELECTOR (current) || select other instrument: OCA
OCA rule set:VIMOS.RLS
Type of rule set:DFOS_OPS
Sub-cascades: ALL       IMG | MOS | IFU
General comment:

Mouse over ...
- a RAW_TYPE to check out the classification rule (M marks a multiple definition)
- a grouping rule to see the complete grouping rule
- a matching rule (e.g. '1,1') to see the product matching rule.
Product tags etc. might be cut off, in that case mouse over the cell to see the full entry.

bottom | right left     
RAW_TYPEACQ_MOS BIAS MOS_CAL M MOS_STDCAL M STD_MOS SCI_MOS
DO_CLASSACQ_MOS BIAS MOS_ARC_SPECTRUM,
MOS_SCREEN_FLAT
MOS_ARC_SPECTRUM,
MOS_SCREEN_FLAT
MOS_STANDARD MOS_SCIENCE
CATGACQUISITION CALIB CALIB CALIB CALIB SCIENCE
Grouping rule single TPL_D TPL_D TPL_D single TPL_D
Index123456

Calibration cascade:
static calibs:master calibs:
2 LINE_CATALOG     1,1 1,1    
5 CONFIG_TABLE     1,1 1,1 1,1 1,1
6 EXTINCT_TABLE         1,1 1,1
9 MOS_MASTER_RESPONSE           0,1
10 STD_FLUX_TABLE         1,1  
    12 MASTER_BIAS 1,1 1,1 1,1 1,1
      13 MOS_ARC_SPECTRUM_EXT...     1,1 MASSOC
      14 MOS_CURV_COEFF   1,1 1,1
      15 MOS_DISP_COEFF   1,1 1,1
      16 MOS_MASTER_SCREEN_FL...     1,1
      17 MOS_SLIT_LOCATION   1,1 1,1
        18 MOS_ARC_SPECTRUM_EXT...   1,1 MASSOC
        19 MOS_CURV_COEFF 1,1 1,1
        20 MOS_DISP_COEFF 1,1 1,1
        21 MOS_MASTER_SCREEN_FL...   1,1
        22 MOS_SLIT_LOCATION 1,1 1,1
          23 MOS_SPECPHOT_TABLE 1,1 MASSOC
          24 MOS_STANDARD_REDUCED... 1,1 MASSOC

Comments:
             

Analysis (technical consistency checks):
products* (CALIB only) n/a OK OK OK OK n/a
mcalibs** n/a n/a OK OK OK OK
powered by QC [createCalibMap v1.3.1] last update: 2013-10-29T10:51:37
*products: does raw_type have a product? (CALIB only)
**mcalibs: are all required mcalibs, per raw_type, somewhere listed as products?
top