NAME QCsinfo_pupil.py
VERSION 1.0 -- 2006-06-27 fill template by handbook
2.0 -- 2008.07.23 completely re-written in Python with added QC functions and images
INSTRUMENT SINFONI
SYNTAX Python
RAWTYPE PUPIL
CALL measureQuality
from $DFS_PRODUCT/PUPIL/$DATE:
QCsinfo_pupil.py -a $AB_name -i $level_of_interaction: 0-2
PURPOSE a) generates QC report, b) writes QC1 parameters into local database
PROCINPUT a) $DATE in now read from AB
b) primary file is set in QCsinfo_pupil.py and is the PRO.CATG=MASTER_PUPIL product with the _0000.fits extension.
c) raw frames are implicitly read from the AB.
d) Reference MASTER_PUPIL is read in from $DFO_QC_DIR/references/PUPIL/.
QC1TABLE trending | table(s) in QC1 database:
sinfoni_pupil
TRENDPLOT trending | current trending plot(s) associated to this procedure:
trend_report_PUPIL_cent_HC.html
QC1PAGE trending | associated trending page(s):
pupil_QC1.html
QC1PLOTS

pupil1.png
display MED_COADD_OBJ pupil on RealTimeDisplay
pupil2.png
UL: Current pupil image. White dotted line denotes pupil centroid.
UC: Reference pupil image.
UR: (reference - current) pupil image.
LL: Difference between pupil cube plane 100 and pupil cube plane 2118.
LR: Filtered current pupil with guassian fit, centroid, and center values listed in legend.
QC1PARAM QC1 parameters, read from the RealTimeComputer of the AO module
AOS.RTC.GUIDESTAR.MAGNITUDE , none
AOS.RTC.OBS.FWHMMEAN , none
AOS.RTC.OBS.FWHMSTD , none
AOS.RTC.OBS.ENCMEAN , none
AOS.RTC.OBS.ENCSTD , none
AOS.RTC.OBS.STREHLMEAN , none

NONE , qc_oes1, oddeven column stripe of first SKY frame, from post-pipeline QC script

QC1 parameters, generated by a post-pipeline QC script that does a Gaussian fit to all planes of the final cube:
X_POS , none
Y_POS , none
FWHM_MIN , none
FWHM_MAX , none
FWHM_POSANG , none
FWHM_RATIO , none
TOT , none

QC1 parameters (a standard sequence)
QC.FRMON.MEANFLUX, qc_frmon_meanflux , standard QC1 parameter: mean flux of lamp=on frame
QC.FRMOFF.MEANFLUX, qc_frmoff_meanflux, standard QC1 parameter: mean flux of lamp=off frame
QC.FRMDIF.MEANFLUX, qc_frmdif_meanflux standard QC1 parameter: mean flux of lamp=on frame minus lamp=off frame
QC.FRMON.MAXFLUX, qc_frmon_maxflux standart QC1 parameter: max flux of lamp=on frame
QC.FRMOFF.MAXFLUX, qc_frmoff_maxflux standart QC1 parameter: max flux of lamp=off frame
QC.FRMDIF.MAXFLUX, qc_frmdif_maxflux standart QC1 parameter: max flux of lamp=on farme minus lam=off frame
QC.FRMON.NPIXSAT , qc_frmon_npixsat Number of saturated pixels

QC1 parameters for PUPIL only
QC.PUPIL.X , qc_pupil_x, (20.207021), pupil center QC.PUPIL.Y , qc_pupil_y (1.741316), pupil center
ALGORITHM Description of algorithms:
qc_pupil_x : centroid position, (weighted average)
qc_pupil_y : centrid position (weighted average)
CERTIF Reasons for rejection:
- As a technical calibration, all position values represent the instrumental/optical status.
- Lamp failure might be a reason the reject the calibration.
COMMENTS None