Next: A modern two-channel photometer
Up: Sample instrument files
Previous: A simple photometer
Next, consider a more modern photometer that uses pulse-counting.
It has a ``neutral'' filter in addition to standard uvby filters.
The ``neutral'' filter is in series with the passband filters, and is used
to determine the dead-time correction.
The positions of these two filter mechanisms are recorded as two adjacent
digits in the data, so we can treat them as a single filter code; the second
digit is 1 when the ``neutral'' filter is in the beam.
There is also a shutter whose position is encoded separately in the data
stream: 0 means open, 1 means shut.
This can be treated as a second logical filter mechanism.
The photomultiplier in this modern instrument is in a thermoelectrically cooled
chamber, regulated to run at
C (273 K).
Table I.14 shows what we get.
Again, there is only one detector, so we need not include the NDETUSED column.
Table:
Instrument table file for a two-channel photometer
Descriptor values:
NFILTCAR = 2
NDETS = 2
FILTSTAT = MEASURED
FILTCAT = ' ' (blank string)
CONDITION = AVERAGE
Passband sub-table
NBAND |
BAND |
FILTCODE_1 |
FILTCODE_2 |
REDLEAK |
NDETUSED |
--- |
-- |
----- |
----- |
---- |
---- |
1 |
U |
1 |
any |
BLOCKED |
1 |
2 |
B |
2 |
any |
BLOCKED |
1 |
3 |
V |
3 |
any |
ABSENT |
1 |
4 |
R |
any |
1 |
BLOCKED |
2 |
5 |
I |
any |
2 |
ABSENT |
2 |
6 |
DARK1 |
0 |
any |
|
1 |
7 |
DARK2 |
any |
0 |
|
2 |
Detector sub-table
NDET |
DET |
SNUMBER |
MODE |
COOLING |
-- |
-- |
---- |
-- |
---- |
1 |
PMT |
S-13 |
CI |
MEASURED |
2 |
PMT |
GAAS |
CI |
DRYICE |
|
Next: A modern two-channel photometer
Up: Sample instrument files
Previous: A simple photometer
http://www.eso.org/midas/midas-support.html
1999-06-15