Xenon-Argon List for ISAAC
--------------------------
Notes
-----
i) Blended at LR means that the line is blended with some other line
at a resolution of 3000. This corresponds to the resolution of the LR
grating with the 0.3" slit. Spectra taken with wider slits will have 
many more blened lines.
ii) Blended at MR means that the lines is blended with some other line
at a resolution of 10000. This corresponds to the resolution of the LR
grating with the 0.3" slit. Spectra taken with wider slits will have 
many more blended lines.
iii) Intensities were calculated from spectra taken with the LR grism
at the standard settings. The standard LR settings are: 0.90, 1.06,
1.25, 1.65 and 2.20 microns.  When lines appear in more than one
setting, the intensity is reported for both settings.
iv) Settings at 0.90, 1.06, 1.25 and 1.65 microns suffer from order
overlap. The lines from different orders are reported in the last
table.
v) Some lines that lie outside the normal LR range are also reported.
vi) The follwoing table lists the details of the LR dispersion solution. 
The residuals can be reduced by using only the brightest lines and
avoiding blends.
                   
Order FWR         EWR           OSF  Arc    NOL  Dispersion  Residual 
---------------------------------------------------------------------
6     7850-10300  8300-9700     Z    Xenon  9    2.38        0.06
                                     Argon  12   2.38        0.12
			
5     9230-12150  9800-11200    SZ   Xenon  19   2.85        0.14
                                     Argon  20   2.85        0.21
			
4    10780-14420  11000-13500   J    Xenon  20   3.56        0.23
                                     Argon  45   3.56        0.24
			
3    14210-19060  14200-18200  SH    Xenon  15   4.74        0.25
                                     Argon  30   4.74        0.19
			
2    18550-25850  19200-25500  SK    Xenon  17   7.13        0.23
                                     Argon  26   7.14        0.37
--------------------------------------------------------------------
OSF = Order sorting filter
FWR = Full wavelength range
EWR = Effective wavelength range.Includes filter, atmosphere and order overlap.
NOL = number of lines used in the fit.
vii) There are a few unidentified lines in the ISAAC spectra, but most
are fainter than 1000 units. 
References
----------
1 - Wavelengths and Transition Probabilities for Atoms and Atomic Ions, 
NSRDS-NBS 68, Reader, Corliss, Wiese and Martin. 
Converted from air to vacuum wavelengths.
2 - CASPIR User's Manual, P. McGregor (MSSSO).
Converted from air to vacuum
3 - CRSP User's Manual (KPNO) - Already in vacuum wavelengths
4 High resolution Argon Atlas by Lloyd Wallace, Ken Hinkle and Dick Joyce 
(KPNO). Already in vacuum wavelengths
See: http://www.noao.edu/kpno/phoenix/tharne/
Conversion from air to vaccuum wavelengths was done via the table in
Astrophysical Quantities, Allen, 1973.
Wavelength	Line	Source	Strength    Comments
--------------------------------------------------------
8266.79         ArI	1	1890
8282.39         XeI	1	1850
8349.11         XeI	1        540
8410.52	      	ArI	1      27660
8426.96	      	ArI	1      49500
8523.78	      	ArI	1      27100
8578.36         XeI	1         50
8608.14	      	ArI	1         50
8650.91         XeI	1        100
8670.32	      	ArI	1       4720 
8694.59         XeI	1         --    Blend at LR 
8699.25         XeI	1         --    Blend at LR
8741.79         XeI	1        320
8760.60         XeI	1         70
8821.83         XeI	1      90480 
8864.75         XeI	1        560
8911.17         XeI	1        260
8933.28         XeI	1        360
8954.71         XeI	1      16350
8983.51         XeI	1         --    Blend at LR 
8990.04         XeI	1         --    Blend at LR 
9047.93         XeI	1      24120
9125.47	      	ArI	1    >140000    Saturated. Too bright to use.
9165.16         XeI	1      31220
9170.03         XeI	1         --    Blend at LR
9197.16         ArI	1        340
9227.03         ArI	1      33200
9294.08         ArI	1        320
9309.19         XeI	2        170
9356.78	      	ArI	1       2080 
9377.33         XeI	1        780
9414.59         XeI	2        180
9447.93         XeI	2        280
9499.67         XeI	2         --    Blend at LR
9515.99         XeI	1       1390
9587.77         XeI	2         30
9660.43         ArI	1      35900 (Z) 11440 (SZ) Overlap b/n Z-SZ
9687.97         XeI	2         40 (Z)   160 (SZ) Overlap b/n Z-SZ
9703.65         XeI	2         20
9720.82         XeI	1        230
9787.18	      	ArI	1      12340
9802.38         XeI	1     120100    
9925.91         XeI	1    >140000    Saturated. Too bright to use
10026.46        XeI	2        480
10054.81      	ArI	1        300
10087.56        XeI	2        210
10110.13        XeI	2       1990
10128.23        XeI	2        150
10166.2	      	Ar	3         40
10191.16        XeI	2         90
10253.84        XeI	2        410
10311.96        ArI	2         50
10335.55        ArI	1        200
10472.92        ArI	1      47810
10480.90        ArI	1         --    Blended at LR
10487.68        XeI	2         80
10509.38        ArI	1        150    Reference 2 gives 10509.35
10530.74        XeI	2       1880
10676.49        ArI	1      11160
10684.70        ArI	1         --    Blended at LR
10703.9         Ar	3        130
10709.71        XeI	2        510
10736.81        ArI	1         70
10761.84        XeI	2        250 
10762.10        ArI	1         80 
10776.30        ArI 	2        120    Overlap with J starts.
10841.34        Xe1	1      16390 (SZ)  1910 (J) Ref. 2 gives 10841.30
10883.94      	ArI 	2        490 (SZ)   130 (J)
10898.31        XeI	2        630 (SZ)   200 (J)
10953.74      	ArI 	2        550 (SZ)   250 (J)
11081.90        ArI	1        930 (SZ)  1090 (J)
11088.27        XeI	2       1120 (SZ)  1410 (J)
11109.50      	ArI 	2        110 (SZ)   150 (J) Uncertain ID
11130.2         XeI	2        230 (SZ)   360 (J)
11165.8         XeI	2         40 (SZ)    40 (J)
11312.7         XeI	2         25 (SZ)    40 (J)
11396.82        Ar	4        470 (SZ)  1450 (J)
11418.2         XeI	2         50 (SZ)   120 (J)
11444.97        ArI	4        180 (SZ) 11450 (J)
11470.69        ArI	4         40 (SZ)   690 (J)
11491.25        ArI	4        380 (SZ) 15000 (J) End of SZ overlap
11617.3	      	Xe	3        420   Beginning of Js
11671.90        ArI	4       8220 
11722.70        ArI	4       2170 
11736.44        ArI	4         --   Blended at LR
11745.5	      	Xe	3       3360
11796.8	      	Xe	3        700
11860.5	      	Xe	3        760
11887.74        ArI	4         --   Blended at LR
11899.89        ArI	4         --   Blended at LR
11915.5	      	XeI	2        140
11946.55        ArI	4       1580   Blended at MR, ArI 11947.14
11956.4	      	Xe	3        340
12029.94        ArI	4       1123
12088.1	      	Xe	3        460
12115.64        ArI	4      17600
12143.06        ArI	4      11310
12154.69        ArI	4         --   Blended at LR
12207.15      	XeI	2         60
12238.59      	XeI	1       1320
12261.11      	XeI	1        220
12346.77        ArI	4        120
12359.67        ArI	4       9210
12406.22        ArI	4      12406
12422.79        ArI	4         --   Blended at LR
12442.73        ArI	4      57730
12459.52        ArI	4      12460
12491.08        ArI	4      12490
12557.76        ArI	4        600
12593.64        XeI	1        560
12599.66        ArI	4        120
12626.84        XeI	1      14150
12641.94        ArI	4        320
12705.75        ArI	4      17640
12736.90        ArI	4       5500
12749.72        ArI	4       4000
12806.24        ArI	4      18230 
12936.73        ArI	4       8500
12960.20        ArI	4      67320
13011.82        ArI	4      15880
13031.99        ArI	4        850
13217.61        ArI	4      27410
13231.72        ArI	4      25290
13234.51        ArI	4         --   Blended at LR
13276.27        ArI	4      39640
13305.95        Ar	4         --   Blended at LR
13316.85        ArI	4      29210	
13333.76        ArI	4       1310
13370.77        ArI	4      61870
13410.26        ArI	4       5230
13503.10        ArI	4         --   Blended at LR
13507.88        ArI	4      91140
13547.91        ArI	4       1450
13577.33        ArI	4       2740
13603.05        ArI	4       3480
13626.38        ArI	4      25460
13660.79      	XeI     1       6460
13682.29        ArI	4      16020   End of Js 
13722.3         Ar      IRAF  107900
13829.5         Ar      IRAF      --   Blended at LR
13832.1         Ar      IRAF      --   Blended at LR
13914.4         Ar      IRAF     630
14097.49        ArI	4        140 (J) Beginning of SH
14146.31      	XeI	1         --   Beyond the normal LR range
14178.58        ArI	4         --   Beyond the normal LR range
14244.85      	XeI	1       1790
14253.09        ArI	4         --   Blended at LR
14260.76        ArI	4         --   Blended at LR, and at MR 14258.04 ArI
14368.91      	XeI	1       1070
14581.44        ArI	4         --   Blended at LR
14600.46        ArI	4       2210   Blended at MR with 14599.73 ArI
14638.42        ArI	4         --   Blended at LR
14654.35        ArI	4         --   Blended at LR
14664.81      	XeI	1        360
14697.03      	ArI	4        910   Blended at MR with 14697.69 ArI
14736.83      	XeI	1      40000
14743.17      	ArI	4       1700
14790.11      	ArI	4        350   ID uncertain
14880.57      	ArI	4         80
15034.62      	ArI	4         --   Blended at LR, and at MR 15035.28 ArI
15050.62      	ArI	4      28600
15103.84      	XeI	1        280
15176.84      	ArI	4       6320
15306.07      	ArI	4       7720
15333.54      	ArI	4       4800
15353.44      	ArI	4         --   Blended at LR
15357.32      	ArI	4         --   Blended at LR
15406.85      	ArI	4       1570
15422.60      	XeI	1      17110
15450.95      	ArI	4        100
15495.20      	XeI	2        110	
15559.71      	ArI	4        180   Blended at MR with XeI 15561.38
15561.38      	XeI	1        420   Blended at MR with ArI 15559.71
15739.21      	ArI	4        200
15797.47      	ArI	4        100
15821.10      	ArI	4        230
15887.51      	ArI	4         --   Blended at LR
15904.03      	ArI	4       2500
15983.90      	XeI	1       1060
15993.86      	ArI	4       8700
16044.28      	XeI	1         --   Blended at LR
16057.66      	XeI	1       6530
16127.06      	ArI	4       1280
16184.44      	ArI	4       1600
16268.51      	ArI	4        250
16441.07      	ArI	4       3140
16524.38      	ArI	4      10400
16553.78      	ArI	4      16553   Blended at MR with 16554.91 ArI
16559.01      	XeI	1        360
16732.72      	XeI	1      11520
16744.65      	ArI	4       3450
16839.14      	XeI	2         60
16887.68      	XeI	2         90
16945.21      	ArI	4      16945
17330.50      	XeI	2       6350 
17369.82      	XeI	2         70
17406.67      	ArI	4        190
17449.67      	ArI	4       4900
17594.8	      	XeI	2        170
17828.87      	ArI	1        890
17919.61      	ArI	4      23690   End of SH
18353.01      	ArI	4         --   Beyond the normal LR range
18423.09      	ArI	4         --   Beyond the normal LR range
18432.80      	ArI	4         --   Blended with 18434.81, 18433.43 ArI
18491.76      	ArI	4         --   Beyond the normal LR range
18569.30      	ArI	4         --   Blended at LR
18575.23      	ArI	4         --   Blended at LR, + MR 18576.65 ArI
18637.37      	ArI	4         330
18750.12      	ArI	4         230
18793.26      	XeI	1        2020
19129.04      	ArI	4	   30
19180.51      	ArI	4          90
19822.91      	ArI	4        3340
19950.52      	ArI	4         --   Blended at LR Check ID and Wav.
19971.18      	ArI	4       3900
20031.14      	ArI	4         --   Blended at LR
20035.57      	ArI	4         --   Blended at LR
20074.41      	ArI	4        240
20192.70      	XeI	1        840
20267.77      	XeI	1      25930
20322.56      	ArI	4       1990
20518.4	      	XeI	2         50
20574.43      	ArI	4        450
20598.9	      	XeI	2        140
20621.86      	ArI	4      31650
20652.77      	ArI	4        890
20722.00      	ArI	4         --   Blended at LR
20739.22      	ArI	4       1200   Blended with 20740.10
20816.72      	ArI	4        820
20890.3	      	XeI	2         50
20991.84      	ArI	4      22440
21041.57      	ArI	4        140
21116.7	      	XeI	2         80
21338.71      	ArI	4       2500
21378.90      	XeI	2        660
21475.94      	XeI	1       4640
21540.09      	ArI	4      10150
22045.58      	ArI	4       3500
22083.21      	ArI	4      12942
22118.66      	ArI	4        360
22275.91      	XeI	2        210
22412.93      	XeI	2        810
22539.74      	ArI	2         40
22624.45      	XeI	2        390
22884.1	      	XeI	2         80
22971.04      	XeI	2        160
22978.37      	ArI	4        150
23028.70      	XeI	2        100
23079.75      	XeI	2        240
23111.56      	XeI	2         50
23139.52      	ArI	4      11910
23199.65      	XeI	1      18000
23285.89      	XeI	1        700
23450.3	      	XeI	2        160
23802.95      	XeI	2        330
23851.54      	ArI	4      10480
23973.06      	ArI	4      13070
24450.4	      	XeI	3       1450
24709.05      	XeI	2        490
24783.37      	ArI	4        150
24831.48      	XeI	1      44600
25132.13      	ArI	4       5760
25152.69      	XeI	1       2500
25166.3	      	XeI	3         --   Blended at LR
25512.18      	ArI	4       3750
25668.02      	ArI	4        900
26276.24      	XeI	1         --   Beyond the normal LR range
---------------------------------------------------------------------
	  
Main lines of other orders.
==============================================
Line     Lamp OSF Order Location Intensity
-----------------------------------------
 9447.93 Xe   Z   5      7873       760
 9515.99 Xe   Z   5      7930      3560
 9660.43 Ar   Z   5      8050     42020
 8283.39 Xe   Z   7      9664       370
 8349.11 Xe   Z   7      9740       200
 8411    Xe   Z   7      9816       400
 8410.52 Ar   Z   7      9812      8700
 8426.96 Ar   Z   7      9831     14703
 8523.78 Ar   Z   7      9944      6490
 8670.32 Ar   Z   7     10115       860
 9660.43 Ar  SZ   6     11593      3420
 9787.97 Ar  SZ   6     11745      3410
 9802.38 Xe  SZ   6     11763     36925
 9925.38 Xe  SZ   6     11910     40624
11081.90 Ar   J   5     13852       700
11396.82 Ar   J   5     14246       660     
11444.97 Ar   J   5     14306       780
11470.69 Ar   J   5     14338       280
11491.25 Ar   J   5     14364      4810
11088.27 Xe   J   5     13860       880
11130.2  Xe   J   5     13913       200
14097.49 Ar  SH   4     18797      6700
14253.09 Ar  SH   4     19004      1860 Blended with 14253.09 Ar
14260.76 Ar  SH   4     19014      1860 Blended with 14260.76 Ar
----------------------------------------------------------------