VIMOS trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2018-04-04T03:30:16 (UT)
  | now: 2024-11-24T16:15:37 (UT) muc03 QC pipeline: vimos-2.9.19 (installed 2016-03-17)
same group: median_master ron_master rms_master ron_raw median_vs_temp   averages_low averages_high
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
VIMOS news:  
Report news:
DATE*: [?] 2018-03-28
no data | NLT
2018-03-29
no data | NLT
2018-03-30
no data | NLT
2018-03-31
no data | NLT
2018-04-01
no data | NLT
2018-04-02
no data | NLT
2018-04-03
no data | NLT
r...
Raw CAL displays:
no CAL raw files no CAL raw files no CAL raw files no CAL raw files no CAL raw files no CAL raw files no CAL raw files
P...
CAL product quality:
no CAL products no CAL products no CAL products no CAL products no CAL products no CAL products no CAL products
 
transfer   ngas [?]

scores&comments | history ... | plot tutorial ... | contact | HC_docu...                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 1.19 ADU VAL1.0,1.4 85 noise_master this | last_yr | all Q2, ron master, low gain
2 QC1DB MEDIAN 1.2 ADU VAL1,1.4 85 noise_master this | last_yr | all Q1, ron master, low gain
3 QC1DB MEDIAN 3.09 ADU VAL2.9,3.6 87 noise_master this | last_yr | all Q2, ron master, high gain
4 QC1DB MEDIAN 3.19 ADU VAL2.9,3.6 87 noise_master this | last_yr | all Q1, ron master, high gain
5 QC1DB MEDIAN 1.13 ADU VAL1.0,1.4 85 noise_master this | last_yr | all Q3, ron master, low gain
6 QC1DB MEDIAN 1.15 ADU VAL0.9,1.3 85 noise_master this | last_yr | all Q4, ron master, low gain
7 QC1DB MEDIAN 3.02 ADU VAL2.8,3.5 87 noise_master this | last_yr | all Q3, ron master, high gain
8 QC1DB MEDIAN 3.11 ADU VAL2.8,3.5 87 noise_master this | last_yr | all Q4, ron master, high gain
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 6|
data source:vimos_bias
(QC1 database)
dataset:noise_master
median:1.19ADU
fixed thresholds:1.0...1.4ADU
N_data plotted:85
[click plot for closeup]
Plot 2
scores: 6|
data source:vimos_bias
(QC1 database)
dataset:noise_master
median:1.2ADU
fixed thresholds:1...1.4ADU
N_data plotted:85
[click plot for closeup]
Plot 3
scores: 6|
data source:vimos_bias
(QC1 database)
dataset:noise_master
median:3.09ADU
fixed thresholds:2.9...3.6ADU
N_data plotted:87
[click plot for closeup]
Plot 4
scores: 6|
data source:vimos_bias
(QC1 database)
dataset:noise_master
median:3.19ADU
fixed thresholds:2.9...3.6ADU
N_data plotted:87
[click plot for closeup]
Plot 5
scores: 6|
data source:vimos_bias
(QC1 database)
dataset:noise_master
median:1.13ADU
fixed thresholds:1.0...1.4ADU
N_data plotted:85
[click plot for closeup]
Plot 6
scores: 6|
data source:vimos_bias
(QC1 database)
dataset:noise_master
median:1.15ADU
fixed thresholds:0.9...1.3ADU
N_data plotted:85
[click plot for closeup]
Plot 7
scores: 6|
data source:vimos_bias
(QC1 database)
dataset:noise_master
median:3.02ADU
fixed thresholds:2.8...3.5ADU
N_data plotted:87
[click plot for closeup]
Plot 8
scores: 6|
data source:vimos_bias
(QC1 database)
dataset:noise_master
median:3.11ADU
fixed thresholds:2.8...3.5ADU
N_data plotted:87
[click plot for closeup]
This plot

Trending plots for bias QC parameters.

All bias parameters are measured in the central 1600x1800 pixels of the detectors. The median bias level and the rms of the master should show a stable trend without sudden jumps or increase/decrease. The read-out noise (ron_raw) is calculated from the difference of the first two raw frames. Some outliers with high values are occasionally observed due to cosmic ray events. The read-out noise as measured from master frames (ron_master) should be extremely stable.

The new detectors (since July 2010) show a correlation of the bias level with temperature, especially for quadrant 2.

Bias frames are usually measured daily during day-time. They come in two different varieties: with low gain (used for imaging) and with high gain (used to calibrate the spectroscopic modes).

Template IDs: img_cal_Dark and spec_cal_Dark. Pipeline recipe: vmbias.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up