[ ESO ]
MIDI:
Quality Control parameters and trending

CAL | HC | refs | QC
QUALITY CONTROL
    HOME
MIDI QC
Trending & QC1
QC1 trending
Pipeline
Data packages
Data Management
QC links:
QC1 LINKS QC1 PLOTS
CURRENT
Detector
Optical elements
Reference Beams
RON | Linearity
Transmission |  Wavelength Calibration
Reference beams: HIGH_SENS | SCI_PHOT | OPEN

Find here detailed information about the MIDI QC1 parameters and select plots. ASCII data from the QC1 database will be available in the future.
GENERAL INFORMATION


Some MIDI calibration data are presently monitored to extract information about their quality and the instrument performance. The quality information is extracted into so-called QC parameters.

All measured QC1 parameters are stored in a database. Their evolution with time is evaluated by the trending process.

 

INSTRUMENT PROPERTIES TRENDED

The following parameters are monitored and used for instrument trending. The plots are available on the MIDI HC page.

detector alignment dispersive elements (GRISM or PRISM) Instrumental Transfer Function
  • More information under the specific TF page

 
[ESO][Index][Search][Help][News]