NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
2022-10-02 The kappa matrix has a validity of 30 days, moreover kappa matrix in MED tilted by 10-20 px does not affect the night observations. The IOT is monitoring the kappa so please ignore for the time being the flag, unless you see for several days in a raw more than 100 px shift. CPa [matisse@eso.org]
2022-12-13 templates are interrupted after only 1/2 the frames, red scores are not significant in this cases [qc_matisse@eso.org]
2022-02-20 Bug in the software. MATISSE Team is aware of this. Working on it.
2023-02-26: the issue seems solved : this was not an pipeline issue but a SHIFT MAP not updated after the switch of lamps (IPE) [ccid@eso.org]
2021-09-28 the maximum OPD is calculated on the 6 baselines. Specifications mentioned a maximum shift of 100 microns/8h for MATISSE-LM and 300 for MATISSE-N. F [qc_matisse@eso.org]