MATISSE trending system: overview of scores

Last update: 2024-11-21T18:24:11 (UT)

General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
MATISSE news:

MATISSE instrument score:   scores only ... | edit COMMENTS ...
HC navigation group group
score
  report scores and comments (all reports)
KPI     no scoring
   
detector_monitoring     no scoring
     
shift     no scoring
   
kappa_matrix   RATIO FLUX SHIFT ZOOM delta_inclination
    2022-10-02 The kappa matrix has a validity of 30 days, moreover kappa matrix in MED tilted by 10-20 px does not affect the night observations. The IOT is monitoring the kappa so please ignore for the time being the flag, unless you see for several days in a raw more than 100 px shift. CPa [matisse@eso.org]
Internal_Vis   VIS2_LM-LOW N CP_LM N DP_LM N SNR_LM N
    2022-12-13 templates are interrupted after only 1/2 the frames, red scores are not significant in this cases [qc_matisse@eso.org] 2022-02-20 Bug in the software. MATISSE Team is aware of this. Working on it.
2023-02-26: the issue seems solved : this was not an pipeline issue but a SHIFT MAP not updated after the switch of lamps (IPE) [ccid@eso.org]
Internal_Flux   FLUX_LM N FLUX_LM_BCDin N_BCDin
   
Internal_OPD   MAX_OPD OPD_LM OPD_N
    2021-09-28 the maximum OPD is calculated on the 6 baselines. Specifications mentioned a maximum shift of 100 microns/8h for MATISSE-LM and 300 for MATISSE-N. F [qc_matisse@eso.org]
Flux_beam   LM-LOW LM-MED L-HIGH L-HIGH+ M-HIGH+ N-LOW N-HIGH LM-LOW_IN-IN N-LOW_IN-IN
   
OPD   Max_LM_OO Max_N_OO Max_LM_II Max_N_II OPD_LM OPD_N
    2021-09-28 For night time observations the OPD must be stable to around 20 microns (for 1% accuracy) [qc_matisse@eso.org]
Transfer_Function   LM-LOW LM-MED L-HIGH L-HIGH+ M-HIGH+ N-LOW N-HIGH
   
Visibility   LM-LOW LM-MED L-HIGH L-HIGH+ M-HIGH+ N-LOW N-HIGH
   
Vis_SNR   LM-LOW LM-MED L-HIGH L-HIGH+ M-HIGH+ N-LOW N-HIGH
   
Closure_Phase   LM-LOW LM-MED L-HIGH L-HIGH+ M-HIGH+ N-LOW N-HIGH
   
Differential_Phase   LM-LOW LM-MED L-HIGH L-HIGH+ M-HIGH+ N-LOW N-HIGH
   
LM_Night_LOW_MED   VIS_LM_LOW MED TF_LOW MED SNR_LOW MED CP_LOW MED DP_LOW MED
   
L_Night_HIGH   VIS_L_HIGH TF_HIGH SNR_HIGH CP_HIGH DP_HIGH
   
L_M_Night_HIGHP   VIS_L_HIGH+ TF_HIGH+ SNR_HIGH+ CP_HIGH+ DP_HIGH+ VIS_M_HIGH+ TF_HIGH+ SNR_HIGH+ CP_HIGH+ DP_HIGH+
   
N_Night_LOW_HIGH   VIS_N_LOW TF_LOW SNR_LOW CP_LOW DP_LOW
   
Correlated_Flux   N_LOW_SNR_CorrFlux CP DP N_HIGH_SNR_CorrFlux CP DP
   
Corr_Fl_Night_range   N_LOW_CorrFlux TF SNR CP DP
   
System_Efficiency   TRANS_LM-LOW LM-MED L-HIGH L-HIGH+ M-HIGH+ N-LOW N-HIGH Corr_AT Corr_UT