LSS: wavelength calibration
old CCD
(before 2007-03-25)
Other HC:
QC links:
[page auto-refreshes after 60 sec]            [press Ctrl+R to enforce refresh of scores and dates]
FORS1 trending system: History Plots
Last update: 2009-04-09T06:55:39 (UT)
| now: 2024-11-22T01:48:28 (UT)  
same group: lambda_c lambda_temp resolution nwave
less ... | history ... | plot tutorial ... | contact                 daily     [?]
Print: Information and research: News:
- png file - advanced studies: QC1 browser | QC1 plotter
 
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB no MEDIAN 5866.96533 Å VAL5867.3,5868.7 82 central_wlen this | last_yr | all C1
2 QC1DB no MEDIAN 5874.45166 Å VAL5875.5,5876.9 82 central_wlen this | last_yr | all C2
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: to indicate that OPSLOG data are included
This plot

After the wavelength calibration is obtained from an LSS arc lamp exposure, the wavelength corresponding to the central CCD pixel is calculated. The values trended here are for the 300V grism with no filter and a slit width of 0.29".  


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up