mirror sites: PL (internal link) HQ     [?]

Health Check monitor

 

CAL | HC | refs | QC
HealthCheck Monitor
   HOME | UsersGuide
  sl icon ALL INSTRUMENTS
 
all KPI reports
ERIS:
dark
Other HC:
QC links:
        [page auto-refreshes after 300 sec]   [stop | on]            [press Ctrl+Shift+R to enforce refresh of scores, dates and news]
ERIS trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-11-13T18:22:46 (UT)
  | now: 2024-11-13T22:34:09 (UT) muc04 QC pipeline: eris-1.6.3 (installed 2024-08-30)
same group: SLOW_dit_1.9 5 60 300 FAST_dit_0.033 0.1 0.15 0.5 2 5   dark_vs_dit
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
ERIS news:  
Report news:
DATE*: [?] 2024-11-06
report | NLT
2024-11-07
report | NLT
2024-11-08
report | NLT
2024-11-09
report | NLT
2024-11-10
report | NLT
2024-11-11
report | NLT
2024-11-12
report | NLT
2024-11-13
no data | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw raw no CAL raw files
P...
CAL product quality:
products products products products products products products no CAL products
 
transfer   ngas [?]

scores&comments | FULL | history ... | plot tutorial ... | contact | HC_docu... | quality events...                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 6.22 ADU VAL2.19,11.23 53 qc_dark_mean this | last_yr | all Clean mean of master dark level
2 QC1DB MEDIAN 3.81 ADU VAL1.74,5.97 53 qc_dark_rms this | last_yr | all Clean standard deviation of master dark
3 QC1DB MEDIAN 1.96e+05 number VAL157600,216500 53 qc_dark_nhotpix this | last_yr | all number of bad pixels
4 QC1DB MEDIAN 0.0468 number VAL0.0376,0.0516 53 qc_dark_bpixfrac this | last_yr | all fration of bad pixels
5 QC1DB MEDIAN 1.78 ADU VAL1.626,1.958 53 qc_dark_readnoise_adu this | last_yr | all detector read out noise (ADU)
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 8|
data source:eris_nix_dark
(QC1 database)
dataset:qc_dark_mean
median:6.22ADU
fixed thresholds:2.19...11.23ADU
N_data plotted:53
[click plot for closeup]
Plot 2
scores: 8|
data source:eris_nix_dark
(QC1 database)
dataset:qc_dark_rms
median:3.81ADU
fixed thresholds:1.74...5.97ADU
N_data plotted:53
[click plot for closeup]
Plot 3
scores: 8|
data source:eris_nix_dark
(QC1 database)
dataset:qc_dark_nhotpix
median:1.96e+05number
fixed thresholds:157600...216500number
N_data plotted:53
[click plot for closeup]
Plot 4
scores: 8|
data source:eris_nix_dark
(QC1 database)
dataset:qc_dark_bpixfrac
median:0.0468number
fixed thresholds:0.0376...0.0516number
N_data plotted:53
[click plot for closeup]
Plot 5
scores: 8|
data source:eris_nix_dark
(QC1 database)
dataset:qc_dark_readnoise_adu
median:1.78ADU
fixed thresholds:1.626...1.958ADU
N_data plotted:53
[click plot for closeup]
This plot

The main qc parameters are plotted for the mostly used DITs  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up