CRIRES trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2014-07-22T11:34:12 (UT)
  | now: 2024-11-21T23:28:02 (UT) muc01 QC pipeline: crire-2.3.2 (installed 2014-03-18)
same group: xcorr ThAr N2O DISP RESOL   AVG
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
CRIRES news:
Report news:
DATE*: [?] 2024-11-14
report | NLT
2024-11-15
report | NLT
2024-11-16
report | NLT
2024-11-17
report | NLT
2024-11-18
report | NLT
2024-11-19
report | NLT
2024-11-20
report | NLT
2024-11-21
no data | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw raw no CAL raw files
P...
CAL product quality:
products products products products products products products no CAL products
  no OPSLOG data  

scores&comments | FULL | history ... | plot tutorial ... | contact                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
Print: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
OPS?
**
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB no MEDIAN 9.54e+04 none VAL75000,120000 89 resolution this | last_yr | all resolving power on chip2
2 QC1DB no MEDIAN 9.91e+04 none VAL75000,125000 89 resolution this | last_yr | all resolving power on chip3
*Data sources: QC1DB: QC1 database; LOCAL: local text file | **OPS: OPSLOG data are included
Plot 1
scores: 8|
data source:crires_wave
(QC1 database)
dataset:resolution
median:9.54e+04none
fixed thresholds:75000...120000none
N_data plotted:89
[click on plot for closeup]
Plot 2
scores: 8|
data source:crires_wave
(QC1 database)
dataset:resolution
median:9.91e+04none
fixed thresholds:75000...125000none
N_data plotted:89
[click on plot for closeup]
This plot

Wavelength calibration: trending of resolving power.

The resolvin power (lambda / Delta lambda) is monitored for the daily health check setting with reference wavelength (on chip 3) of 1090.4 nm which uses a ThAr lamp. It is determined from the median of the spectral lines identified on each detector.  


General information

Click on any of the plots to see a close-up version. The latest date is indicated on top of the plot, data points belonging to that date are specially marked.

If configured,

  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up