CRIRES trending system: HEALTH CHECK report
HELP USERS-GUIDE MORE
Last update: 2024-11-21T16:22:37 (UT)
  | now: 2024-11-21T19:34:16 (UT) muc03 QC pipeline: cr2re-1.6.2 (installed 2024-10-18)
same group: DARK_med_YJHK DARK_med_LM DARK_RON DARK_med DARK_nbad
General news: NOTE: The Health Check and calChecker services are moved to qcFlow on an instrument by instrument basis. The current pages will not be updated any longer once the move has been finished.
CRIRES news:
Report news:
DATE*: [?] 2024-11-14
report | NLT
2024-11-15
report | NLT
2024-11-16
report | NLT
2024-11-17
report | NLT
2024-11-18
report | NLT
2024-11-19
report | NLT
2024-11-20
report | NLT
r...
Raw CAL displays:
raw raw raw raw raw raw raw
P...
CAL product quality:
products products products products products products products
 
transfer     [?]

scores&comments | history ... | plot tutorial ... | contact | HC_docu... | quality events...                 daily/often; important to check     [?]
*Date on this monitor changes at 21:00 UT
png: Information and research:
- png file - advanced studies: QC1 browser | QC1 plotter
Plot
?
Symb
?
Source
*
Average ? Thresholds ? N_
data
QC1
parameter
Data
downloads
Remarks
method value unit method value
1 QC1DB MEDIAN 4.61e+04 NONE VAL40000,50000 88 dark_nbad this | last_yr | all DARK_nbad, wlen_id=J1228, dit=3.0, CHIP1
2 QC1DB MEDIAN 5.56e+04 NONE VAL49000,59000 88 dark_nbad this | last_yr | all DARK_nbad, wlen_id=J1228, dit=3.0, CHIP2
3 QC1DB MEDIAN 8.26e+04 NONE VAL75000,85000 88 dark_nbad this | last_yr | all DARK_nbad, wlen_id=J1228, dit=3.0, CHIP3
4 QC1DB MEDIAN 35 K VAL34.95,35.05 1591 det_temp this | last_yr | all detector temperature, INS.TEMP29.VAL
5 QC1DB MEDIAN 6.74e+04 NONE VAL62000,72000 87 dark_nbad this | last_yr | all DARK_nbad, wlen_id=J1228, dit=90.0, CHIP1
6 QC1DB MEDIAN 8.24e+04 NONE VAL77000,87000 87 dark_nbad this | last_yr | all DARK_nbad, wlen_id=J1228, dit=90.0, CHIP2
7 QC1DB MEDIAN 1.19e+05 NONE VAL114000,124000 87 dark_nbad this | last_yr | all DARK_nbad, wlen_id=J1228, dit=90.0, CHIP3
*Data sources: QC1DB: QC1 database; LOCAL: local data source
Plot 1
scores: 8|
data source:criresp_dark
(QC1 database)
dataset:dark_nbad
median:4.61e+04NONE
fixed thresholds:40000...50000NONE
N_data plotted:88
[click plot for closeup]
Plot 2
scores: 8|
data source:criresp_dark
(QC1 database)
dataset:dark_nbad
median:5.56e+04NONE
fixed thresholds:49000...59000NONE
N_data plotted:88
[click plot for closeup]
Plot 3
scores: 8|
data source:criresp_dark
(QC1 database)
dataset:dark_nbad
median:8.26e+04NONE
fixed thresholds:75000...85000NONE
N_data plotted:88
[click plot for closeup]
Plot 5
scores: 7|
data source:criresp_dark
(QC1 database)
dataset:dark_nbad
median:6.74e+04NONE
fixed thresholds:62000...72000NONE
N_data plotted:87
[click plot for closeup]
Plot 6
scores: 7|
data source:criresp_dark
(QC1 database)
dataset:dark_nbad
median:8.24e+04NONE
fixed thresholds:77000...87000NONE
N_data plotted:87
[click plot for closeup]
Plot 7
scores: 7|
data source:criresp_dark
(QC1 database)
dataset:dark_nbad
median:1.19e+05NONE
fixed thresholds:114000...124000NONE
N_data plotted:87
[click plot for closeup]
Plot 4
scores: no current data
data source:criresp_dark
(QC1 database)
dataset:det_temp
median:35K
fixed thresholds:34.95...35.05K
N_data plotted:1591
[click plot for closeup]
This plot

Trending plot for the number of bad pixels in the CRIRES master DARK calibrations taken in the J1228 wavelength setting, for each of the detectors.

The number of bad pixels in the darks INS.WLEN.ID=J1228, short DET.DIT=3.0s (2.0s until 2023-04-05) are shown in the upper sub-plots, and for the long DET.DIT=90.0s in the lower sub-plots.
Sub-plot number 4 shows detectors' temperature, read from the INS.TEMP29.VAL keyword.  


General information

Click on any of the plots to see a close-up version.

The latest date is indicated on top of the plot.

If configured,

  • data points belonging to the latest date are specially marked
  • statistical averages are indicated by a solid line, and thresholds by broken lines
  • outliers are marked by a red asterisk. They are defined as data points outside the threshold lines
  • "aliens" (= data points outside the plot Y limits) are marked by a red arrow ( or )
  • you can download the data for each parameter set if the 'Data downloads' link shows up